Skip to the content

FDA Medical Device Recalls (openFDA)

Oxus America, Inc.: Li-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400. Also branded as Reliability Plus byDrive. The Portable Oxygen Concentrator is intended to deliver concentrated oxygen for adult patients with chronic pulmonary diseases.

Recall number
Z-0768-2013
Recalling firm
Oxus America, Inc.
Product
Li-Ion battery used in the Oxus Portable Oxygen Concentrator (POC), RS-00400. Also branded as Reliability Plus byDrive. The Portable Oxygen Concentrator is intended to deliver concentrated oxygen for adult patients with chronic pulmonary diseases.
Status
Terminated
Initiated
2013-01-14
Posted
2013-02-02
Terminated
2013-07-17
Reason
Due to past battery inventory storage conditions and deficiencies in the maintenance procedure, it was possible that some batteries were over-discharged to point of damaging the cells.
Root cause
Nonconforming Material/Component
Action
Oxus, Inc. sent an "URGENT: MEDICAL DEVICE RECALL" letter dated January 11, 2013 to all affected customers.The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-888-475-1568 for information related to this recall.
Quantity
1724 total units
Distribution
Worldwide Distribution-USA including the states of New York, Michigan, Colorado, Texas, Iowa, Florida, Georgia, Ohio, Wisconsin, Oregon, Alabama, Illinois, Utah, Oklahoma, West Virginia and California and the countries of Canada, European Union, South Africa, Korea and Brazil.
Lot, serial or model codes
part number RS-00402 Unit ID # (TS-): 001418, 001419, 001420, 001421, 001422, 001423, 001424, 001425, 001426, 001427, 001428, 001429, 001430, 001431, 001432, 001433, 001434, 001435, 001436, 001437, 001438, 001439, 001440, 001441, 001442, 001443, 001444, 001445, 001446, 001447, 001448, 001449, 001450, 001451, 001452, 001453, 001454, 001455, 001456, 001457, 001458, 001459, 001460, 001461, 001462, 001463, 001464, 001465, 001466, 001467, 001468, 001469, 001470, 001471, 001472, 001473, 001474, 001475, 001476, 001477, 001478, 001479, 001480, 001481, 001482, 001483, 001484, 001485, 001486, 001487, 001488, 001489, 001490, 001491, 001492, 001493, 001494, 001495, 001496, 001497, 001498, 001499, 001551, 001552, 001557, 001558, 001559, 001560, 001561, and 001562.    Battery SN (MB-): 06625x, 06666x, 06679x, 06715x, 08335x, 06498x, 01578x, 12009x, 12010x, 00075x, 05624x, 11868x, 11855x, 11587x, 11601x, 08240x, 08336x, 06145x, 06160x, 06302x, 06305x, 08189x, 08190x, 09200x, 09562x, 09564x, 09654x, 09719x, 09720x, 09787x, 09788x, 09887x, 09902x, 09903x, 09904x, 09905x, 09906x, 09957x, 09958x, 09959x, 09960x, 09961x, 10049x, 10477x, 10478x, 10479x, 10480x, 05689x, 05690x, 05694x, 05695x, 05696x, 05727x, 07937x, 07938x, 07940x, 07941x, 08072x, 08112x, 08125x, 10358x, 11165x, 11166x, 11961x, 11027x, 08628x, 08630x, 09143x, 09144x, 09145x, 09146x, 09498x, 09888x, 09889x, 09890x, 09891x, 09983x, 09984x, 10547x, 10549x, 10550x, 01476x, 01527x, 01576x, 05630x, 08657x, 08677x, 08752x, 08753x, 08755x, 08838x, 09724x, 10237x, 10582x, 10597x, 10600x, 11252x, 11253x, 11264x, 11265x, 11266x, 11271x, 11272x, 11469x, 11472x, 11473x, 11525x, 11526x, 11527x, 11528x, 11999x, 01518x, 09212x, 10584x, 10602x, 11110x, 11426x, 11427x, 11456x, 11471x, 11478x, 11485x, 11486x, 11514x, 11518x, 11589x, 11591x, 11592x, 11593x, 05647x, 05648x, 05649x, 06646x, 06672x, 06673x, 06674x, 06675x, 06676x, 06702x, 07558x, 07637x, 07638x, 07639x, 07640x, 07641x, 07643x, 07644x, 07645x, 07663x, 07680x, 07689x, 07691x, 07697x, 07703x, 07704x, 07705x, 08257x, 08258x, 08259x, 08260x, 08649x, 08650x, 08681x, 08807x, 08808x, 08811x, 09304x, 09387x, 09388x, 09390x, 09859x, 09860x, 10089x, 10402x, 10613x, 10614x, 10616x, 05407x, 06072x, 06078x, 06080x, 06081x, 06209x, 06217x, 06226x, 06277x, 06278x, 06279x, 06299x, 06329x, 06331x, 06396x, 06432x, 06433x, 06434x, 06436x, 06442x, 06443x, 06444x, 06445x, 06446x, 06474x, 06475x, 06476x, 06481x, 06512x, 06515x, 06517x, 06518x, 06522x, 06525x, 06810x, 06872x, 06874x, 06877x, 07030x, 07112x, 07113x, 07114x, 07116x, 07154x, 07169x, 07170x, 07171x, 07173x, 07194x, 05589x, and 07749x.
510(k) numbers
["K020324"]
PMA numbers
not on file
Product code
CAW
Firm FEI
3008841036
Firm city
Rochester Hills
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register