FDA Medical Device Recalls (openFDA)

Smiths Medical ASD, Inc.: TTS-400J Level1 Tympanic Temperature Sensor with 400 Series Thermistor

Recall number
Z-0768-06
Recalling firm
Smiths Medical ASD, Inc.
Product
TTS-400J Level1 Tympanic Temperature Sensor with 400 Series Thermistor
Status
Terminated
Initiated
2006-03-08
Posted
2006-04-18
Terminated
2012-05-30
Reason
Product temperature readings may be 2C higher than the actual temperature
Root cause
Other
Action
Smiths Medical notified consignees by letter dated 3/8/05. Customers are instructed to complete customer reply notices and return affected products for replacement.
Quantity
400
Distribution
Nationwide, Canada, France, Japan, Netherlands
Lot, serial or model codes
Lot Number: 3080
510(k) numbers
["K864195"]
PMA numbers
not on file
Product code
FLL
Firm FEI
1221261
Firm city
Rockland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register