Getinge Usa Sales Inc: Maquet Prismalix OR Light Systems, Model Numbers and Catalog Numbers 3001 - ARD567424001C, 3001 HMS07 - ARD567722001C, 3001 HXS12 - ARD567703001C, 4001 S/DF - ARD567211211C, 4001 SAD/DF - ARD567221211C, 4001 SAD/DF - ARD567221241C, 4001 SAD/SF - ARD567221141C, 4001 SAI/DF - ARD567220111C, 4001K S/DF - ARD567211213C, 4003 - ARD567215111C, 4401 S/DF - ARD567212211C, 4401 SAD/DF - ARD567223211C, 4401 SAD/DF - ARD567223241C, 4401 SAD/SF - ARD567223111C, 4441 SAD/DF - ARD567225241C, 6001K ACS/T - ARD567234113C, 6001K SAL/T - ARD567226113C, 6101 ACS/SU - ARD567235001C, 6301 SAL/T - ARD567423111C, 6401 ACS/DF - ARD567235211C, 6401 ACS/DF - ARD567235231C, 6401 ACS/DF - ARD567235241C, 6401 SAL/DF - ARD567227211C, 6401 SAL/DF - ARD567227241C, 6401K ACS/D - ARD567235243C, 6401K SAL/SF - ARD567227113C, 6411 SAL/DF - ARD567228009C, 6441 ACS/DF - ARD567236241C, 8401 ACS/DF - ARD567231225C, 8401 ACS/DF - ARD567231241C, 8401K ACS/D - ARD567231243C, 8431 ACS/DF - ARD567431221C, 8431 ACS/DF - ARD567431241C, 8441 ACS/DF - ARD567232241C, PRISMALIX - ARD567761211C

Recall number
Z-0767-2024
Recalling firm
Getinge Usa Sales Inc
Product
Maquet Prismalix OR Light Systems, Model Numbers and Catalog Numbers 3001 - ARD567424001C, 3001 HMS07 - ARD567722001C, 3001 HXS12 - ARD567703001C, 4001 S/DF - ARD567211211C, 4001 SAD/DF - ARD567221211C, 4001 SAD/DF - ARD567221241C, 4001 SAD/SF - ARD567221141C, 4001 SAI/DF - ARD567220111C, 4001K S/DF - ARD567211213C, 4003 - ARD567215111C, 4401 S/DF - ARD567212211C, 4401 SAD/DF - ARD567223211C, 4401 SAD/DF - ARD567223241C, 4401 SAD/SF - ARD567223111C, 4441 SAD/DF - ARD567225241C, 6001K ACS/T - ARD567234113C, 6001K SAL/T - ARD567226113C, 6101 ACS/SU - ARD567235001C, 6301 SAL/T - ARD567423111C, 6401 ACS/DF - ARD567235211C, 6401 ACS/DF - ARD567235231C, 6401 ACS/DF - ARD567235241C, 6401 SAL/DF - ARD567227211C, 6401 SAL/DF - ARD567227241C, 6401K ACS/D - ARD567235243C, 6401K SAL/SF - ARD567227113C, 6411 SAL/DF - ARD567228009C, 6441 ACS/DF - ARD567236241C, 8401 ACS/DF - ARD567231225C, 8401 ACS/DF - ARD567231241C, 8401K ACS/D - ARD567231243C, 8431 ACS/DF - ARD567431221C, 8431 ACS/DF - ARD567431241C, 8441 ACS/DF - ARD567232241C, PRISMALIX - ARD567761211C
Status
Open, Classified
Initiated
2023-11-08
Posted
2024-01-19
Terminated
not on file
Reason
A potential for a light system to fall in the operating room.
Root cause
Equipment maintenance
Action
An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing MSA808092.sw@getinge.com Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to or-lights2023.qrc@getinge.com or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin
Quantity
236793 units in total
Distribution
US Nationwide. Global Distribution.
Lot, serial or model codes
All systems affected
510(k) numbers
["K982063"]
PMA numbers
not on file
Product code
FSY
Firm FEI
3013876692
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register