Skip to the content

FDA Medical Device Recalls (openFDA)

WEST PHARMA. SERVICES IL, LTD: Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only

Recall number
Z-0767-2019
Recalling firm
WEST PHARMA. SERVICES IL, LTD
Product
Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only
Status
Terminated
Initiated
2018-12-26
Posted
2019-01-28
Terminated
2022-07-20
Reason
Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.
Root cause
Device Design
Action
On December 26, 2018, West Pharmaceutical Services, Inc. issued URGENT MEDICAL DEVICE CORRECTION and ACKNOWLEDGEMENT FORMS to their customers asking customers to avoid use of the drug oxytocin with the device. Actions to be taken by the Distributor: 1. Complete the required Acknowledgement and Receipt Form below and return (via email) to Kevin.Lentz@westpharma.com when complete (See attached form). 2. Prior to filling order / releasing shipments of any Vial2Bag DC 13mm products (which have been placed on hold per our request) to hospital customers, affix a copy of this Voluntary Field Corrective Action Notice to each carton of Vial2Bag DC 13mm products in your possession. 3. Notify and provide the letter to all accounts/customers where the Affected Product was distributed. 4. Because the assessment is ongoing, please be prepared for updates to the notification. Actions to be taken by the Customer/User: 1. As a temporary measure you should avoid use of the drug oxytocin with the Vial2Bag DC 13mm until further notice until a full assessment has been completed. 2. Affix the notice to each carton of Vial2Bag DC 13mm products in your possession. 3. In addition, provide the notice to personnel with responsibilities related to product selection and administration, and according to your internal procedures for notification of field corrective actions. 4. Please complete the Recall Acknowledgement and Receipt Form and return to Karen Crimi by email at cs@progressivemedinc.com when complete responses are available. On January 24, 2019, West Pharmaceutical Services, Inc. initiated a product removal and broadened their action by distributing AMENDED URGENT MEDICAL DEVICE RECALL notices to their customers to include removal of three (3) devices from the field. On February 1, 2019, West Pharmaceutical Services, Inc. expanded their recall to include an additional sixteen (16) lots and issued notices titled - 2nd AMENDED URGENT MEDICAL DEVICE RECALL not
Quantity
3.9 million units
Distribution
US Distributor in MO
Lot, serial or model codes
UDI:  (01)07290108240061(17)211031(10)A839(90)3609812 5(91)6070111  Lot Numbers: 7153, 7154, 7634, 7688, 7752, 7909, 7955, 8152, 8153, 8154, 8232, 8233, 8234, 8235, 9028, 9175, 9029, 9176, 9177, 9178, 9367, 9368, 9369, 9370, 9654, 9656, 9655, 9918, 9710, 9711, 9712, 9713, 9839, 9840, 9983, 9984, 9992, 9993, 9994, 9995, A123, A124, A790, A805, A806, A791, A792, A793, A794, A795, A796, A797, A807, A808, A809, A810, A834, A835, A836, A837, A838 & A839
510(k) numbers
["K170095"]
PMA numbers
not on file
Product code
LHI
Firm FEI
3000223297
Firm city
Ra'Anana
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register