Villa Sistemi Medicali S.P.A.: Apollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.

Recall number
Z-0765-2020
Recalling firm
Villa Sistemi Medicali S.P.A.
Product
Apollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
Status
Open, Classified
Initiated
2019-11-04
Posted
not on file
Terminated
not on file
Reason
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Root cause
Device Design
Action
Urgent Field Safety Notice notification letters dated 10/23/19 were sent to customers. Action to be taken by the customer/user: The user shall take note of amendment/reinforcement of Instructions for Use (IFU) and Service Manual, attached to this FN and, if it is the case, request a service technician to do the periodic preventive maintenance the user shall replace the attached pages in the User Manual and in the Service Manual, prescribing the periodic replacement of the angulation chain. Fill the table present in the addendum; In the cell "Next replacement date" write the date adding 8 years to the replacement date. All the examinations performed on the device before the recall remain valid. The IFU and the Service Manual need to be amended to prescribe the replacement of the angulation chain after 8 years of service life of the unit. This notice needs to be passed to all users of Apollo, Apollo DRF, Apollo EZ and to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred.
Quantity
7
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
Lot, serial or model codes
Serial Numbers: 08090531 09090591 10040680 15051450 11020798 17011602 17041639
510(k) numbers
["K050190"]
PMA numbers
not on file
Product code
KXJ
Firm FEI
3002808461
Firm city
Buccinasco
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register