FDA Medical Device Recalls (openFDA)
Villa Sistemi Medicali S.P.A.: Apollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
- Recall number
- Z-0765-2020
- Recalling firm
- Villa Sistemi Medicali S.P.A.
- Product
- Apollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
- Status
- Open, Classified
- Initiated
- 2019-11-04
- Posted
- not on file
- Terminated
- not on file
- Reason
- An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- Root cause
- Device Design
- Action
- Urgent Field Safety Notice notification letters dated 10/23/19 were sent to customers. Action to be taken by the customer/user: The user shall take note of amendment/reinforcement of Instructions for Use (IFU) and Service Manual, attached to this FN and, if it is the case, request a service technician to do the periodic preventive maintenance the user shall replace the attached pages in the User Manual and in the Service Manual, prescribing the periodic replacement of the angulation chain. Fill the table present in the addendum; In the cell "Next replacement date" write the date adding 8 years to the replacement date. All the examinations performed on the device before the recall remain valid. The IFU and the Service Manual need to be amended to prescribe the replacement of the angulation chain after 8 years of service life of the unit. This notice needs to be passed to all users of Apollo, Apollo DRF, Apollo EZ and to all those who need to be aware within your organisation or to any organisation where the potentially affected devices have been transferred.
- Quantity
- 7
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
- Lot, serial or model codes
- Serial Numbers: 08090531 09090591 10040680 15051450 11020798 17011602 17041639
- 510(k) numbers
- ["K050190"]
- PMA numbers
- not on file
- Product code
- KXJ
- Firm FEI
- 3002808461
- Firm city
- Buccinasco
- Firm state
- not on file
- Firm country
- Italy
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28