FDA Medical Device Recalls (openFDA)
Philips Electronics North America Corporation: Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.
- Recall number
- Z-0765-2016
- Recalling firm
- Philips Electronics North America Corporation
- Product
- Allura Xper FD OR Table, Ceiling version only; 722035; imaging applications.
- Status
- Terminated
- Initiated
- 2015-11-03
- Posted
- 2016-02-05
- Terminated
- 2016-08-19
- Reason
- Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced position, the air gap of the brake can become too large, resulting in continued C-arc movement (C-arc is looking for balanced position).
- Root cause
- Device Design
- Action
- The letters for the customers in the United States were sent on 11/3/2015, via certified return receipt and were addressed to the Risk Manager. Philips will perform a test on the 15 affected Allura Xper R8.2 FD 20C systems to see if the motion controller amplifier of the brake motor works correctly. If the test will not pass, the motor will be replaced. The Recall Package is sent to all Key Markets worldwide. The Key Market of an affected country will contact the facility where these systems were installed.
- Quantity
- 2 subject to correction and removal
- Distribution
- Worldwide Distribution. US states of TN and IN and the countries of Australia, Belgium, Egypt, France, Germany, Japan, Republic of Korea, Netherlands, and Saudi Arabia.
- Lot, serial or model codes
- Ceiling version only.
- 510(k) numbers
- ["K102005"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 1218950
- Firm city
- Andover
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28