MEDLINE INDUSTRIES, LP - Northfield: Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223

Recall number
Z-0764-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Status
Open, Classified
Initiated
2024-12-16
Posted
2025-01-15
Terminated
not on file
Reason
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Root cause
Under Investigation by firm
Action
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 12/16/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-270-FG Recall Code: 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. Update: A follow-up letter titled IMMEDIATE ACTION REQUIRED dated 2/11/25 was sent to customers. Medline would like to update our action to instruct you to please destroy all impacted convenience kits and Medline will provide replacement kit(s) with the correct Fluid Delivery Set. REQUIRED ACTION: 1. Immediately check your stock for the affected kit number(s) and lot number(s) listed on the attachment. Please destroy all affected kits. 2. Please complete and return the enclosed destruction form listing the quantity of affected kits destroyed. Even if you do not have any affected product, please complete and return the form. 3. Upon receipt of your completed destruction form, Medline will issue replacem
Quantity
13,600 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;
510(k) numbers
["K852140"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register