Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: WECK Visistat 35W, Disposable Skin Stapler, Rx Only, Sterile, Teleflex Medical, NC 27709. The intended use of the Visistat stapler is to close lacerations on the outer layer of the dermis for a wide range of general, thoracic, obstetric, gynecological, orthopedic, cardiovascular, urological, and plastic surgical procedures.

Recall number
Z-0764-2017
Recalling firm
Teleflex Medical
Product
WECK Visistat 35W, Disposable Skin Stapler, Rx Only, Sterile, Teleflex Medical, NC 27709. The intended use of the Visistat stapler is to close lacerations on the outer layer of the dermis for a wide range of general, thoracic, obstetric, gynecological, orthopedic, cardiovascular, urological, and plastic surgical procedures.
Status
Terminated
Initiated
2016-11-18
Posted
not on file
Terminated
2017-11-24
Reason
Sterility Deficiency: The packages are not sealed properly and thus the sterility is not assured.
Root cause
Packaging process control
Action
The firm, Telexflex Medical, sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letter dated 11/18/2016 to its Consignees/Customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately discontinue use and quarantine any affected products; complete and return the Recall Acknowledgement Form by fax to 1-855-419-8507, Attn: Customer Service or email to: recalls@teleflex.com (even if you do not have any affected product) and return the product with the completed Recall Acknowledgement Form. A customer service representative will contact you with a RGA # and provide instructions for the return of the product. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Quantity
96,402 units (9,120 ea US, 87,282 ea OUS)
Distribution
Worldwide Distribution- U.S. (Nationwide) and countries of: Australia, Belgium, Canada, China, Germany, Hong Kong, India, Japan, Korea, Malaysia, New Zealand, Singapore and Thailand.
Lot, serial or model codes
Product Code: 528235, Lot numbers: 73C1600693, 73G1500681, 73H1500255, 73H1500256, 73K1500618, 73L1400006 and 73M1500130.
510(k) numbers
not on file
PMA numbers
not on file
Product code
GDT
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.