Getinge Usa Sales Inc: Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C

Recall number
Z-0759-2024
Recalling firm
Getinge Usa Sales Inc
Product
Maquet Lucea - Lucea10/40, Lucea50/100 OR Light Systems, Model Numbers and Catalog Numbers LCA 10 - ARD2LCA00040C, LCA 10 - ARDLCA109000C, LCA 10 - ARDLCA109002C, LCA 10 - ARDLCA109006C, LCA 10 - ARDLCA209008A, LCA 10 - ARDLCA209008C, LCA 10 - ARDLCA209009A, LCA 10 - ARDLCA209009C, LCA 10 - ARDLCA209010A, LCA 10 - ARDLCA209010C, LCA 10 - ARDLCA409000C, LCA 10 - ARDLCA409001C, LCA 10 - ARDLCA409002A, LCA 10 - ARDLCA409002C, LCA 10 - ARDLCA409003C, LCA 10 - ARDLCA409004A, LCA 10 - ARDLCA409004C, LCA 10 - ARDLCA409005C, LCA 10 DES - ARDLCA409001A, LCA 10 MOBILE - ARDLCA309008A, LCA 10 RAIL - ARDLCA409005A, LCA 10 WALL - ARDLCA109006A, LCA 40 - ARDLCA109001C, LCA 40 - ARDLCA208000A, LCA 40 - ARDLCA208000C, LCA 40 - ARDLCA208001A, LCA 40 - ARDLCA208001C, LCA 40 - ARDLCA209000A, LCA 40 - ARDLCA209000C, LCA 40 - ARDLCA209001A, LCA 40 - ARDLCA209001C, LCA 40 - ARDLCA209002A, LCA 40 - ARDLCA209002C, LCA 40 - ARDLCA209003A, LCA 40 - ARDLCA209003C, LCA 40 - ARDLCA209014A, LCA 40 - ARDLCA209014C, LCA 40 - ARDLCA209017A, LCA 40 - ARDLCA209017C, LCA 40 - ARDLCA209018A, LCA 40 - ARDLCA209018C, LCA 40 - ARDLCA209019A, LCA 40 - ARDLCA209019C, LCA 40 - ARDLCA209020A, LCA 40 - ARDLCA209020C, LCA 40 - ARDLCA209021A, LCA 40 - ARDLCA209021C, LCA 40 MOBILE - ARDLCA309009A, LCA 40 WALL - ARDLCA109001A, LCA 40GS - ARDLCA109009A, LCA 40GS - ARDLCA109009C, LCA 100 - ARDLCA109005A, LCA 100 - ARDLCA109005C, LCA 100 - ARDLCA219000A, LCA 100 MOBILE - ARDLCA309006A, LCA 100 MOBILE B - ARDLCA309007A, LCA 100 SAT - ARDSAT209004A, LCA 100 SAT - ARDSAT209005A, LCA 100 V - ARDLCA219000C, LCA 100DF - ARDLCA109008A, LCA 100DF - ARDLCA109008C, LCA 100DF - ARDLCA219003A, LCA 100DF - ARDSAT209008A, LCA 100DF - ARDSAT209009A, LCA 100DF V - ARDLCA219003C, LCA 50 - ARDLCA109004A, LCA 50 - ARDLCA109004C, LCA 50 - ARDLCA209012A, LCA 50 - ARDLCA209012C, LCA 50 - ARDSAT209002A, LCA 50 MOBILE - ARDLCA309004A, LCA 50 MOBILE B - ARDLCA309005A, LCA 50 SAT - ARDSAT209003A, LCA 50 WALL 11 - ARD2LCA00140A, LCA 50DF - ARDLCA109007A, LCA 50DF - ARDLCA109007C, LCA 50DF - ARDLCA209015A, LCA 50DF - ARDLCA209015C, LCA 50DF - ARDSAT209006A, LCA 50DF - ARDSAT209007A, LCA DUO 100 - ARDLCA219001A, LCA DUO 100 V - ARDLCA219001C, LCA DUO 100+50 - ARDLCA219002A, LCA DUO 100+50 V - ARDLCA219002C, LCA DUO 100+50DF - ARDLCA219005A, LCA DUO 100+50DF V - ARDLCA219005C, LCA DUO 100DF - ARDLCA219004A, LCA DUO 100DF V - ARDLCA219004C, LCA DUO 50 - ARDLCA209013A, LCA DUO 50 - ARDLCA209013C, LCA DUO 50DF - ARDLCA209016A, LCA DUO 50DF - ARDLCA209016C, LCA100 MOBILE UPGRADE - ARDLCA506006A, LCA100 UPGRADE - ARDLCA506004A, LCA100 UPGRADE 24V - ARDLCA506005A, LCA100 WALL UPGRADE - ARDLCA506007A, LCA50 UPGRADE - ARDLCA506000A, LCA50 UPGRADE 24V - ARDLCA506001A, LCA50 UPGRADE WALL MOUNTED - ARDLCA506003A, SAT LCA 100 - ARDSAT209004C, SAT LCA 100 - ARDSAT209005C, SAT LCA 100DF - ARDSAT209008C, SAT LCA 100DF - ARDSAT209009C, SAT LCA 50 - ARDSAT209002C, SAT LCA 50 - ARDSAT209003C, SAT LCA 50DF - ARDSAT209006C, SAT LCA 50DF - ARDSAT209007C, LCA 10 MOB - ARDLCA309000A, LCA 10 MOB - ARDLCA309000C, LCA 10 MOB - ARDLCA309002A, LCA 10 MOB - ARDLCA309002C, LCA 10 MOB - ARDLCA309008C, LCA 100 MOBILE - ARDLCA309006C, LCA 100 MOBILE B - ARDLCA309007C, LCA 40 MOB - ARDLCA309001A, LCA 40 MOB - ARDLCA309001C, LCA 40 MOB - ARDLCA309003A, LCA 40 MOB - ARDLCA309003C, LCA 40 MOB - ARDLCA309009C, LCA 40GS MOBILE - ARDLCA309010A, LCA 40GS MOBILE - ARDLCA309010C, LCA 50 MOBILE - ARDLCA309004C, LCA 50 MOBILE B - ARDLCA309005C,
Status
Open, Classified
Initiated
2023-11-08
Posted
2024-01-19
Terminated
not on file
Reason
A potential for a light system to fall in the operating room.
Root cause
Equipment maintenance
Action
An URGENT MEDICAL DEVICE CORRECTION letter dated 11/8/23 was sent to customers. Actions to be taken by the customer: Our records indicate that you have been identified as having purchased at least one model of one of these product ranges. Please identify your affected units using the referenced devices on page 1 and follow the instructions below. Please distribute this notice and forward the link below to access manuals to anyone within your organization who may need to use the manuals, and especially to your service or service provider who performs your maintenance. Light system fixing and other replacements: Regarding: - the fixing of the light systems: suspensions fixing screws, adapter fixing screws, bushing fixing screws - the brake screws - the safety segments - the batteries, a dedicated instruction to illustrate and facilitate understanding of what to replace and what to order for replacement has been issued. The instruction is available on a dedicated space on the web portal https://www.getinge.com/int/campaigns/maintenance-and-service-on-or-light-system/, and provided upon request by emailing MSA808092.sw@getinge.com Should you have questions or require additional information, please do not hesitate to contact Maquet SAS/Getinge or your local representative. Please complete and sign the attached MEDICAL DEVICE CORRECTION - RESPONSE FORM (page 7) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to or-lights2023.qrc@getinge.com or by faxing the form to (866) 350-0897. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Actions to be taken by Maquet SAS / Getinge: As the majority of customers may not have regular contact with Getinge services or authorized Getinge representatives and therefore may not be aware of current maintenance recommendations, Maquet SAS/Getinge is providin
Quantity
236793 units in total
Distribution
US Nationwide. Global Distribution.
Lot, serial or model codes
All systems affected UDI-DI: 3700712415921 3700712415853 3700712415846 3700712415877 3700712415938 3700712415860 3700712400750 3700712400613 3700712400712 3700712400729 3700712401115 3700712401122 03700712400767 3700712400620 3700712401139 3700712401146 3700712400774 3700712400637 3700712401153 3700712400736 3700712400743 3700712401160 03700712400781 3700712400644 3700712401177 3700712401184 3700712400651 3700712400668 3700712400675 3700712400682 3700712400699 03700712400705 03700712404710 03700712404697 03700712404703 03700712404727 03700712404666 03700712404673 03700712404680
510(k) numbers
["K113679"]
PMA numbers
not on file
Product code
FTD
Firm FEI
3013876692
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register