GE Medical Systems, LLC: Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).

Recall number
Z-0758-2023
Recalling firm
GE Medical Systems, LLC
Product
Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).
Status
Open, Classified
Initiated
2022-11-10
Posted
2022-12-22
Terminated
not on file
Reason
Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
Root cause
Device Design
Action
GE Healthcare notified consignees on about 11/10/2022 via letter titled "URGENT MEDICAL DEVICE CORRECTION. For U.S. customers, letters were sent via FedEx. Consignees were instructed to follow updated safety instructions provided in the appendix included with the notification letter, place the appendix with the product labeling, and replace the batter every 2 years or if the battery is not capable of powering the system for more than 30 minutes. In response, GE Healthcare requested consignees return the response form.
Quantity
113 units
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, AZERBAIJAN, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, CHINA Colombia, CONGO, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, CZECH REPUBLIC, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, HONG KONG, Hungary, ICELAND, INDIA, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Kuwait, KYRGYZSTAN, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, LUXEMBOURG, Macedonia, MALAYSIA, Mali, Mauritania, Mauritius, Mexico, MICRONESIA, Moldova, Morocco, Mozambique, Namibia, Nepal, Netherlands, NEW ZEALAND, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, SYRIA, TAIWAN, TANZANIA, THAILAND, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Lot, serial or model codes
No US Distribution. Model Number H45561AC, System ID: DZ1237US02, AO1015US01, EG1843US01, EG5882US01, EG1621US04, EG1621US03, 2538978U01, 083038042857318, 083038602007811, 083038602826113, 0835384095, IQ1165US05, IQ1376US01, KE1049US01, KW1010US04, KW1102US04, KW1001US35, KW1011US21, KW1011US22, KW1006US14, KW1098US06, KW1020US15, KT1061US01, KT1062US02, LB1359US04, 1553UL1026, 0834380243, 0834384085, 0834384095, MA1251US13, MA1272US10, OM1042US03, OM1007US11, OM1007US12, PK1331US08, PK1331US06, PA1097US01, SA1201US02, SA1014US07, SA1337US05, SA1060US08, SA1147US05, SA1149US03, SA1051US09, SA2220US02, SA2220US01, SA2072US17, SA1028US06, SA1089US07, SA1138US05, SA2158US07, SA1340US02, SA1223US15, SA1002US08, SA1036US06, SA1025US03, SA2214US03, SA1272US10, SA1024US04, SA1101US04, SA1034US20, SA1249US11, SA1034US23, SA1058US60, SA1058US47, SA1058US59, SA1332US03, SA1030US05, SA1282US05, ZA2573US01, ZA2422US01, ZA2328US01, ZA2339US01, ZA2420US01, ZA1024US07, ZA2348US01, ZA2358US01, ZA1030US10, ZA2421US01, ZA1025US16, ZA1025US25, ZA2378US01, ZA2510US01, TN1028US04, 01002ULT25, 06004ULT19, 07021ULT02, P34189UL17, 21002ULT18, 23002ULT06, P23003UL12, 32004ULT18, 34013ULT13, 34037ULT16, 34013ULT14, P35076UL03, 38006ULT08, 42002ULT12, 42002ULT10, P50005UL02, 55001ULT13, 60001ULT07, P61029UL05, AE1283US01, AE1353US01, AE1016US16, AE1016US14, AE1016US15, AE1011US18, AE1331US01, AE1015US16, YE1071US04, YE1071US03
510(k) numbers
not on file
PMA numbers
not on file
Product code
ITX
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register