Baxter Healthcare Renal Div: Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.

Recall number
Z-0758-04
Recalling firm
Baxter Healthcare Renal Div
Product
Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
Status
Terminated
Initiated
2004-03-10
Posted
2004-07-20
Terminated
2006-10-26
Reason
Microbubbles of air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian.
Root cause
Other
Action
Urgent Device Correction letters dated 3/11/04 were sent to all Meridian users, informing them of adverse events associated with air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian. The letter outlined the clinical practices that must be followed for all hemodialysis therapies to reduce the potential for air to be introduced into the extracorporeal blood circuit during hemodialysis. Any questions or related events to report were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 2. Baxter developed a permanent fix -- an enhanced air detector with improved sensitivity, making it capable of detecting air bubbles of a smaller size. Letters were sent to all Meridian customers on 7/11/05, informing them of the availability of the new detector and that a Baxter Field Service Engineer would contact them to schedule the upgrade.
Quantity
2,949 units
Distribution
Nationwide, Mexico, China and Korea.
Lot, serial or model codes
all serial numbers
510(k) numbers
["K992894"]
PMA numbers
not on file
Product code
KDI
Firm FEI
3000210122
Firm city
Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register