FDA Medical Device Recalls (openFDA)
Baxter Healthcare Renal Div: Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
- Recall number
- Z-0758-04
- Recalling firm
- Baxter Healthcare Renal Div
- Product
- Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085 U.S.A.
- Status
- Terminated
- Initiated
- 2004-03-10
- Posted
- 2004-07-20
- Terminated
- 2006-10-26
- Reason
- Microbubbles of air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian.
- Root cause
- Other
- Action
- Urgent Device Correction letters dated 3/11/04 were sent to all Meridian users, informing them of adverse events associated with air in the extracorporeal blood circuit during hemodialysis therapy on the Meridian. The letter outlined the clinical practices that must be followed for all hemodialysis therapies to reduce the potential for air to be introduced into the extracorporeal blood circuit during hemodialysis. Any questions or related events to report were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 2 and choice 2. Baxter developed a permanent fix -- an enhanced air detector with improved sensitivity, making it capable of detecting air bubbles of a smaller size. Letters were sent to all Meridian customers on 7/11/05, informing them of the availability of the new detector and that a Baxter Field Service Engineer would contact them to schedule the upgrade.
- Quantity
- 2,949 units
- Distribution
- Nationwide, Mexico, China and Korea.
- Lot, serial or model codes
- all serial numbers
- 510(k) numbers
- ["K992894"]
- PMA numbers
- not on file
- Product code
- KDI
- Firm FEI
- 3000210122
- Firm city
- Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28