Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570

Recall number
Z-0757-2021
Recalling firm
Boston Scientific Corporation
Product
Contour VL Variable Length Ureteral Stent, UPN M0061801550, M0061801560, M0061801570
Status
Open, Classified
Initiated
2020-11-16
Posted
not on file
Terminated
not on file
Reason
Trace amounts of metallic debris were found on the stent related to the manufacturing process. The debris was confirmed to be particles of nickel.
Root cause
Process design
Action
On November 16, 2020, the firm distributed Urgent Medical Device Removal letters to affected customers. Customers were informed of the product issue and asked to take the following actions: - Verify whether you have any product from the affected lots/batches within your inventory - Immediately cease use or distribution of any remaining unused product from the affected lots/batches - Segregate affected material and return it to BSC in accordance with removal instructions included in the letter. - Distributors should forward the notification to their customers Your local Sales Representative can answer any questions that you may have regarding this notification.
Quantity
458 units
Distribution
Distribution US nationwide and Japan.
Lot, serial or model codes
GTIN 08714729058410:  Lots  25578577, 25598812, 25609220, 25614542, 25614545, 25619687, 25641771, 25713852, 25753933, 25790349.   GTIN 08714729067986:  Lots  25640715, 25670518, 25686790, 25686791, 25686792, 25741763, 25741770, 25753937, 25790351, 25795704, 25802759, 25809481, 25822104.   GTIN 08714729068020:  Lots  25578740, 25583824.
510(k) numbers
["K974541"]
PMA numbers
not on file
Product code
FAD
Firm FEI
3005099803
Firm city
Marlborough
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register