FDA Medical Device Recalls (openFDA)
Conmed Corporation: PADPRO ADULT Radiotranslucent Pad w/Medtronic connector, REF/Catalog #2516M. RX ONLY. Defibrillation/Pacing/Cardioversion/Monitoring Electrodes. Made in USA for: CONMED CORPORATION Utica, New York, 13502.
- Recall number
- Z-0757-2007
- Recalling firm
- Conmed Corporation
- Product
- PADPRO ADULT Radiotranslucent Pad w/Medtronic connector, REF/Catalog #2516M. RX ONLY. Defibrillation/Pacing/Cardioversion/Monitoring Electrodes. Made in USA for: CONMED CORPORATION Utica, New York, 13502.
- Status
- Terminated
- Initiated
- 2007-03-08
- Posted
- 2007-04-24
- Terminated
- 2012-07-16
- Reason
- Firm was notified of three incidents over the past 15 months in which customers opened the pouch & found an incorrect connector. This error makes the product unusable. The Zoll and Medtronic connectors are not interchangeable and are unique to their specific device.
- Root cause
- Other
- Action
- On 3/08/07, "URGENT DEVICE RECALL" letters (dated 3/05/07) will be sent via UPS to the US distributors and direct user accounts. The letters inform them of the problem and of the recall. The letters instruct them to immediately discontinue using the affected lots, inspect their stocks, and to return them as soon as possible. The letters instruct the consignees to notify their customers of the recall of the devices to the user level if they have been distributed, and request the return of the products. Additionally, when requested, ConMed will notify distributor's customers (sub-accounts) by the same means. Returned devices will be automatically replaced by ConMed Corporation. A separate letter is being sent to the foreign accounts today instructing them to return the devices back to ConMed Corporation for final disposition. The letters also instruct the consignees to complete the accompanying reply form and fax it to Nancy Crisino at (315) 624-3089. ConMed has contracted Stericycle in Indianapolis, IN, to handle the recall and returns from the domestic accounts.
- Quantity
- 920 pairs
- Distribution
- Worldwide, including USA, Canada, Romania, Chile, and Belgium
- Lot, serial or model codes
- Lot Y101105-4
- 510(k) numbers
- ["K020203"]
- PMA numbers
- not on file
- Product code
- MLN
- Firm FEI
- 3001451571
- Firm city
- Utica
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28