Skip to the content

FDA Medical Device Recalls (openFDA)

BioMerieux SA: API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in the involved API strips.

Recall number
Z-0756-2016
Recalling firm
BioMerieux SA
Product
API ZYM Bx2 (Ref 70493) Product Usage: ZYM B reagent is an additional test used for revealing the results of some miniaturized biochemical test including in the involved API strips.
Status
Terminated
Initiated
2016-01-08
Posted
2016-02-04
Terminated
2017-07-14
Reason
Discrepant identification results related to API identification strips. These discrepant identification results are due to false negative results on the tests of the strips linked to the ZYM B reagent.
Root cause
Other
Action
BioMrieux sent an Urgent Field Safety Notice letter DATED jANUARY 8, 2015 to affected customers. The letter identified the affected product, problem and action to be taken. dated . The letter indicated that they could continue to use product in inventory with mandatory requirements that before using a ZYM B ampule and each day of use of API strips that they perform a quality control with ATCC strains as described in each package insert of involved API strips. The letter also requested that they distribute this letter to all appropriate personnel in the laboratory, retain a copy for their files, and forward the information to all parties that may use the product. For additional assistance or have any questions, contact your local bioMrieux Customer Service representative.
Quantity
13,271
Distribution
Nationwide Distribution in the states of: AL, CA, CO, CT, DC, FL, GA, HI, IL, IN, IA, KS, KY, ME, MD, MA, MI, MN, MS, MO, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TX, VT, VA, WA, WV, WI, WY
Lot, serial or model codes
Lot No. PF - 1003765610/Lot No. Ampules - 1003693040/Exp. date 1/15/16 Lot No. PF - 1003772410/Lot No. Ampules - 1003711170/Exp. date 1/19/16 Lot No. PF - 1003772470/Lot No. Ampules - 1003733320/Exp. date 1/29/16 Lot No. PF - 1003787970/Lot No. Ampules - 1003693020/Exp. date 1/12/16 Lot No. PF - 1003790030/Lot No. Ampules - 1003790010/Exp. date 2/26/16 Lot No. PF - 1004077210/Lot No. Ampules - 1004024590/Exp. date 5/25/16 Lot No. PF - 1004097610/Lot No. Ampules - 1004040410/Exp. date 6/2/16 Lot No. PF - 1004355410/Lot No. Ampules - 1004306800/Exp. date 9/20/16 Lot No. PF - 1004355730/Lot No. Ampules - 1004306790/Exp. date 9/20/16
510(k) numbers
not on file
PMA numbers
not on file
Product code
LIB
Firm FEI
1000611339
Firm city
Marcy L'Etoile
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register