FDA Medical Device Recalls (openFDA)
Stryker Endoscopy: PKG, ASSY, 5MM PEEK MULTI HANDLE, 45CM, P/N 0250181108 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
- Recall number
- Z-0756-2015
- Recalling firm
- Stryker Endoscopy
- Product
- PKG, ASSY, 5MM PEEK MULTI HANDLE, 45CM, P/N 0250181108 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
- Status
- Terminated
- Initiated
- 2014-11-17
- Posted
- 2014-12-12
- Terminated
- 2015-07-17
- Reason
- The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
- Root cause
- Error in labeling
- Action
- Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Customers with questions were instructed to contact Stryker Endoscopy at 408-855-6238 or via email at endorecall@stryker.com. For questions regarding this recall call 408-754-2000.
- Quantity
- 180,573 units total
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lot, serial or model codes
- 0200136B 0343780C 0441296C 0443152C 0540019C 0543638D 6112014 7312014 0845029D 0847459D 0847460D 0848127D 0940087D 0941072D 0943442D 0943690D 0944823D 1040626D 1040928D 1041984D 1043027D 1044518D 1045056D 1046104D 1046956D 1140857D 1141971D 1142616D 1143647D 1144959D 1145342D 1145878 1145878D 1146673D 1241510D 1242259D 1243096D 1243551 1243551D 1244182 1244645 1245796 1245797 1245797D 1245978 1245978H 1246407 1340934 1340934H 1341256 1341256H 1344090 1440292 1441860 1442306 9801674S
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- KOG
- Firm FEI
- 2936485
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28