Teleflex Medical: Percuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.

Recall number
Z-0754-2019
Recalling firm
Teleflex Medical
Product
Percuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
Status
Terminated
Initiated
2018-01-03
Posted
not on file
Terminated
2020-10-30
Reason
There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.
Root cause
Process control
Action
The firm, Teleflex, sent an "Urgent Medical Device Recall" letter dated 1/3/2019 to its customers on 01/03/2019. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1.If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com or hand it to your sales representative. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-396-2111.
Quantity
1962 units
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Iran, Ireland, Italy, Japan, Lebanon, Netherlands, Qatar, Romania, Saudi Arabia, Singapore, Spain, Taiwan, Turkey, and the United Arab Emirates.
Lot, serial or model codes
Batch/Lot Numbers: 73A1700535, 73B1600170, 73B1700073, 73B1700523, 73B1700646, 73C1600058, 73C1600706, 73D1600482, 73F1600086, 73G1600207, 73G1600468, 73G1600573, 73G1700041, 73G1700311, 73G1700577, 73H1600509, 73H1700513, 73H1700526, 73H1700531, 73H1700532, 73J1700100, 73J1700553, 73M1600016, 73M1600027, 73M1600231
510(k) numbers
["K153063"]
PMA numbers
not on file
Product code
GEI
Firm FEI
3005747797
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register