FDA Medical Device Recalls (openFDA)
Greatbatch Medical: 45 Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662, T3635, T5586,T6031, T6164,T10161, T12628,T14879, T16422,T2487,T2713,T116057, T13635, MPU260045AO01, MPU260045ZH01. The Driver is intended to act as an interface between the cutting tool and a motorized driving system (i.e. surgical drill).
- Recall number
- Z-0751-2013
- Recalling firm
- Greatbatch Medical
- Product
- 45 Angle Driver / Z-H, Greatbatch Medical 10 000 Wahrie Drive, Clarance, NY 14031. Model #: P2661, P2662, T3635, T5586,T6031, T6164,T10161, T12628,T14879, T16422,T2487,T2713,T116057, T13635, MPU260045AO01, MPU260045ZH01. The Driver is intended to act as an interface between the cutting tool and a motorized driving system (i.e. surgical drill).
- Status
- Terminated
- Initiated
- 2013-01-16
- Posted
- 2013-01-30
- Terminated
- 2014-09-16
- Reason
- Greatbatch Medical has initiated a global voluntary recall and market withdrawal of its entire Small Angled Driver 45 product line. Greatbatch has received reports of noisy and / or rough operation when using these Small Angled Drivers.
- Root cause
- Nonconforming Material/Component
- Action
- Greatbatch sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 15, 2012 to all affected customers. The letter described the Reason For Recall, Risk to Health, and Instructions to Customers. Advised consignees to examine their inventory and quarantine the product in question. If they have further distributed the product to notify their customers and to complete and return the "Recall Response Form". Contact the firm at 1-763-951-8235 for questions relating to this notice.
- Quantity
- 913
- Distribution
- Worldwide Distribution-USA including the states of CA, IN, KS, MA, NJ, PA, TN, and TX, and the countries of AUSTRALIA, BRAZIL, CANADA, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, PORTUGAL, and SWITZERLAND.
- Lot, serial or model codes
- All lots
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 2183787
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28