Roche Molecular Systems, Inc.: LightMix Zika rRT-PCR Test The LightMix Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. 263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.

Recall number
Z-0750-2017
Recalling firm
Roche Molecular Systems, Inc.
Product
LightMix Zika rRT-PCR Test The LightMix Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. 263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.
Status
Terminated
Initiated
2016-11-16
Posted
not on file
Terminated
2017-12-12
Reason
The LightCycler 480 algorithm used for the LightMix Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.
Root cause
Software design
Action
Roche Molecular Diagnostics Inc. sent an Urgent Medical Device Correction Letter dated November 16, 2016, to all affected customers via UPS. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product until Roche provides updated Instructions for Use to assist in result interpretation. If product was further distributed, ensure the Urgent Medical Device Correction letter is provided to that site. Also, customers were instructed to completed the attached fax form and fax it to 1-888-912-2188. For questions customers should call 1-800-526-1247.
Quantity
24 pieces
Distribution
Nationwide Distribution
Lot, serial or model codes
38161611
510(k) numbers
not on file
PMA numbers
not on file
Product code
POA
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register