FDA Medical Device Recalls (openFDA)
Waldemar Link GmbH & Co. KG (Mfg Site): Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05
- Recall number
- Z-0749-2024
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Product
- Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05
- Status
- Open, Classified
- Initiated
- 2023-11-30
- Posted
- 2024-01-18
- Terminated
- not on file
- Reason
- Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time
- Root cause
- Under Investigation by firm
- Action
- Link Bio Corp. issued Urgent Medical Device Recall Letter via email on 11/30/23. Letter states reason for recall, health risk and action to take: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801. Please contact LinkBio at 973-625-1333 or complaints@linkbio.com for any questions.
- Quantity
- 1 unit
- Distribution
- US Nationwide distribution in the state of NH.
- Lot, serial or model codes
- UDI-DI :04026575245253 Lot / Serial Number: C306165
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FZX
- Firm FEI
- 3003386935
- Firm city
- Norderstedt
- Firm state
- not on file
- Firm country
- Germany
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28