Alere San Diego, Inc.: Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL VERS) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), and glucose (GLU) test procedures on the Alere Cholestech LDX System. Calibration verification is optional for CLIA waived systems, such as the Alere Cholestech LDX System, under CLIA regulations. However local or state regulations may require that calibration verification is run at regular intervals.
- Recall number
- Z-0747-2013
- Recalling firm
- Alere San Diego, Inc.
- Product
- Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL VERS) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), and glucose (GLU) test procedures on the Alere Cholestech LDX System. Calibration verification is optional for CLIA waived systems, such as the Alere Cholestech LDX System, under CLIA regulations. However local or state regulations may require that calibration verification is run at regular intervals.
- Status
- Terminated
- Initiated
- 2012-10-22
- Posted
- 2013-01-29
- Terminated
- 2013-09-26
- Reason
- Out of range (increased) HDL Cholesterol results.
- Root cause
- Component design/selection
- Action
- A recall letter was sent on 10/22/12 to all customers who purchased the Alere Cholestech LDX Mutianalyte Control, 2 vial set, Catalog Number 88769, Lot #C2510 & Lot #C2510A. The letter informed the customers that these lots of controls may yield results which are out of range high for the Level 2 control when tested with the Alere Cholestech LDX HDL test. Customers are instructed of the actions to be taken. On 11/16/12, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog 88770, Lot#271097. The problem identified is that these lots of calibration verification materials may yield results which are out of range high when tested with the Alere Cholestech LDX HDL test. Customers are instruted of the actions to be taken. On 3/6/13, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 Vial set, Catalog # 88770, Lot#298148 and 301886. The recall expanded to include the Alere Cholestech LDX Multianalyte Control, 2 vial Set, Catalog # 88769, Lots#C2572, C2572A, C2593 and C2593A. Customers are instructed to complete and fax the enclosed verification form within 10 days. Customers with questions are instructed to contact Alere San Diego, 9975 Summers Ridge Rd, San Diego, CA 92121, (877) 308-8289, fax (858) 805-8457, and responses.ts@alere.com.
- Quantity
- 614 kits
- Distribution
- Worldwide Distribution - USA (nationwide) Puerto Rico and South Korea.
- Lot, serial or model codes
- Lot#271097, 298148, and 301886
- 510(k) numbers
- ["K102700"]
- PMA numbers
- not on file
- Product code
- JJY
- Firm FEI
- 1000125596
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28