Olympus America Inc.: Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended for use in electrosurgery procedures during flexible gastrointestinal endoscopy.
- Recall number
- Z-0747-2012
- Recalling firm
- Olympus America Inc.
- Product
- Olympus Electrosurgical Unit. The Olympus Electrosurgical Unit (ESG-100) is used in conjunction with cutting and coagulation electrosurgical devices intended for use in electrosurgery procedures during flexible gastrointestinal endoscopy.
- Status
- Terminated
- Initiated
- 2011-11-28
- Posted
- 2012-01-13
- Terminated
- 2013-03-01
- Reason
- Investigations of customer complaints revealed a counterfeit capacitor was used in the manufacture of some lots of the ESG-100.
- Root cause
- Nonconforming Material/Component
- Action
- Olympus America Inc. sent an URGENT: MEDICAL DEVICE CORRECTION ACTION" letter dated November 28, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue use of the affected product. A questionnaire was attached to the letter for customers to complete and return via fax at 484-896-7128. Contact the firm at 484-896-5000 for questions regarding this notice.
- Quantity
- 245
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Serial numbers 11053P230001, 11053P230002, 11053P230003, 11053P230004, 11053P230005, 11060P230001, 11160P230002, 11060P230003, 11160P230004, 11160P230005, 11081P230001, 11081P230003, 11081P230004, 11081P230005, 11081P230006, 11081P230007, 11081P230008, 11081P230009, 11081P230010, 11094P230006, 11094P230007, 11094P230008, 11094P230009, 11094P230010, 11096P230007, 11096P230008, 11096P230009, 11096P230010, 11102P230001, 11102P230002, 11102P230003, 11102P230004, 11102P230005, 11109P230002, 11109P230003, 11109P230004, 11109P230005, 11110P230001, 11110P230002, 11110P230003, 11110P230004, 11110P230005, 11116P230001, 11116P230002, 11116P230003, 11116P230004, 11116P230005, 11117P230001, 11117P230002, 11117P230003, 11117P230004, 11117P230005, 11132P230001, 11132P230002, 11132P230003, 11132P230004, 11132P230005, 11133P230001, 11133P230002, 11133P230003, 11138P230002, 11138P230003, 11138P230004, 11139P230001, 11139P230002, 11139P230005, 11167P230001, 11167P230004, 11180P230003, 11180P230005, 11201P230005, 11201P230010, 11202P230001, 11202P230005, 25456W23-101, 25456W23-102, 25456W23-107, 25456W23-108, 25456W23-109, 25522W23-101, 25522W23-102, 25522W23-103, 25522W23-104, 25522W23-105, 25553W23-105, 25553W23-107, 25553W23-108, 25553W23-109, 25553W23-110, 25565W23-102, 25565W23-103, 25565W23-105, 25565W23-106, 25565W23-108, 25565W23-109, 25565W23-110, 25593W23-101, 25593W23-102, 25593W23-103, 25593W23-104, 25593W23-105, 25593W23-107, 25593W23-109, 25593W23-110, 25607W23-101, 25607W23-102, 25607W23-103, 25607W23-104, 25607W23-105, 25607W23-106, 25607W23-107, 25607W23-108, 25607W23-110, 25613W23-101, 25613W23-102, 25613W23-103, 25613W23-104, 25613W23-105, 25613W23-106, 25613W23-107, 25613W23-108, 25613W23-109, 25613W23-110, 25686W23-102, 25719W23-105, 25719W23-106, 25736W23-101, 25736W23-102, 25736W23-103, 25736W23-104, 25736W23-105, 25736W23-106, 25736W23-107, 25736W23-108, 25736W23-109, 25752W23-101, 25752W23-102, 25752W23-103, 25752W23-104, 25752W23-105, 25752W23-106, 25752W23-107, 25776W23-101, 25776W23-102, 25776W23-103, 25776W23-105, 25776W23-106, 25776W23-107, 25776W23-108, 25776W23-109, 25776W23-110, 25869W23-101, 25869W23-102, 25869W23-103, 25869W23-104, 25869W23-105, 25869W23-106, 25869W23-107, 25869W23-108, 25869W23-109, 25869W23-110, 25978W23-104, 25978W23-105, 25978W23-106, 25978W23-107, 26052W23-109, 26052W23-110, 26085W23-101, 26085W23-102, 26085W23-103, 26085W23-104, 26085W23-105, 26085W23-106, 26085W23-107, 26085W23-108, 26085W23-109, 26136W23-101, 26136W23-103, 26136W23-104, 26136W23-105, 26136W23-107, 26136W23-108, 26136W23-109, 26136W23-110, 26162W23-108, 26162W23-109, 26162W23-110, 26227W23-102, 26227W23-103, 26227W23-104, 26227W23-105, 26227W23-106, 26227W23-107, 26227W23-108, 26227W23-109, 26227W23-110, 26456W23-101, 26456W23-102, 26456W23-103, 26456W23-104, 26456W23-105, 26456W23-106, 26456W23-107, 26456W23-108, 26456W23-109, 26456W23-110, 26559W23-101, 26559W23-103, 26559W23-104, 26559W23-105, 26559W23-108, 26559W23-109, 26559W23-110, 26710W23-108, 26710W23-109, 26770W23-101, 26770W23-102, 26770W23-103, 26770W23-104, 26770W23-105, 26770W23-106, 26770W23-107, 26770W23-108, 26770W23-109, 26770W23-110, 26878W23-101, 26878W23-102, 26878W23-103, 26878W23-104, 26878W23-105, 26879W23-102, 26879W23-103, 26879W23-104, 26879W23-105, 26917W23-101, 26917W23-102, 26917W23-103, 26917W23-105, 26917W23-106, 26917W23-107, 26917W23-108, 26917W23-109, 26917W23-110, and 26992W23-110.
- 510(k) numbers
- ["K073207"]
- PMA numbers
- not on file
- Product code
- GEI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28