Skip to the content

FDA Medical Device Recalls (openFDA)

Westmed, Inc.: REF 0562, Adult Elongated O2/CO2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supply Tubing, Rx Only, CE0482, (01)00709078001697 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.

Recall number
Z-0746-2021
Recalling firm
Westmed, Inc.
Product
REF 0562, Adult Elongated O2/CO2 Mask, 10'X0.50 GSL w/ Swivel Elbow, 2" Pigtail w/Female Luer Connectors, 7' Oxygen Supply Tubing, Rx Only, CE0482, (01)00709078001697 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.
Status
Terminated
Initiated
2020-11-11
Posted
not on file
Terminated
2023-02-07
Reason
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Root cause
Process control
Action
On November 11, 2020, West Med, Inc. issued a recall notification for oxygen delivery products via E-Mail to affected consignees. In addition to informing consignees about the recall products, the firm asked consignees to take the following actions: 1. Please complete the attached response form and quarantine affected product. 2. DO NOT DESTROY OR DISPOSE OF PRODUCT AT THIS TIME. 3. Contact Kalyn Galvez-Latneau, Westmed Customer Service Manager, for additional information and to obtain a Return Material Authorization (RMA). Kalyn Galvez-Latneau, kgalvez@westmedinc.com, (520) 294-7987, Ext 1232.
Quantity
6000 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
Lot, serial or model codes
Lot # 090820N30
510(k) numbers
["K913164"]
PMA numbers
not on file
Product code
KGB
Firm FEI
1000374126
Firm city
Tucson
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register