Skip to the content

FDA Medical Device Recalls (openFDA)

Tosoh Smd Inc: AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706

Recall number
Z-0745-2020
Recalling firm
Tosoh Smd Inc
Product
AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706
Status
Terminated
Initiated
2019-11-15
Posted
not on file
Terminated
2020-07-01
Reason
The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898.
Root cause
Device Design
Action
Tosoh Bioscience issued Urgent Medical Device Correction on 11/15/19, letter states reason for recall, health risk and action to take: Complete and return the attached Acknowledgement Form to Tosoh Bioscience, Inc., within 14-days of receiving this notification by FAX, email, or postal mail as indicated on the form. In the event you receive Error Code 2220, discard the affected vial(s). You may pre-screen pretreatment 1 vials on AIA-2000 analyzers prior to reconstituting the vial to help identify vials that may return Error Code 2220. This is an optional pre-screen step and is not required for use of vials from the impacted lots. This is a voluntary customer notification that should be provided to all staff members operating the Tosoh AIA-2000 Analyzer.. Please maintain awareness of this notification by filing it with your laboratory records and forward this information to others who may have received this product. Should you have any questions regarding this medical device notification, please feel free to contact Bernadette OConnell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Quantity
1012 units
Distribution
Nationwide Foreign: Columbia, Dominic Republic, Ecuador
Lot, serial or model codes
Lot Numbers: IZ01893 Exp. 2019-07-31;  I801890 Exp. 2019-11-30
510(k) numbers
["K964348"]
PMA numbers
not on file
Product code
CDD
Firm FEI
3005529799
Firm city
Grove City
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register