FDA Medical Device Recalls (openFDA)
Becton Dickinson & Company: BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.
- Recall number
- Z-0745-2017
- Recalling firm
- Becton Dickinson & Company
- Product
- BD 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing Catalog # 367342 Intended for venipuncture to obtain blood specimens from patients.
- Status
- Open, Classified
- Initiated
- 2016-08-08
- Posted
- not on file
- Terminated
- not on file
- Reason
- BD has received multiple reports of units associated with the 23 G x .75 in. BD Vacutainer Push Button Blood Collection Set with 12 in. Tubing whose lots are exhibiting separation of front and rear barrels upon activation of the safety feature, which retracts the needle.
- Root cause
- Device Design
- Action
- The firm, BD, sent a "URGENT PRODUCT RECALL" letter dated 8/8/2016 to its customer with a return response form via UPS 2nd Day mail on 8/8/2016.The letter described the product, problem and actions to be taken. The customers were instructed to immediately review your inventory for the specific lot numbers, quarantine the product and discontinue the shipment of the affected product; complete and return the Recall Response Form by fax to BD at 1-866-913-0661 or email to bd6835@stericycle.com (even if you do not have any product); return all affected products with completed response form per packing instructions; if you have distributed this product, identify your customers and notify them immediately of this product recall; if you would like BD to conduct the notification to your customers, email your customer list within 3 business days to Matthew.Kelleher@bd.com. if you have any questions or require assistance with the return of the recalled product and/or availability of credit product, contact 1-888-628-0732 between 8Am and 5 PM ET Monday through Friday.
- Quantity
- 1,788,150 units
- Distribution
- Worldwide distribution- US (nationwide) and countries of: Canada, Belgium, Brazil.
- Lot, serial or model codes
- 5355531 5355533 5356681 5356682 6006629 6006630 6006631 6006633 6006634 6006650
- 510(k) numbers
- ["K030573"]
- PMA numbers
- not on file
- Product code
- JKA
- Firm FEI
- 2243072
- Firm city
- Franklin Lakes
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28