Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8161. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.

Recall number
Z-0744-2012
Recalling firm
Baxter Healthcare Corp.
Product
Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8161. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.
Status
Terminated
Initiated
2010-07-14
Posted
2012-01-13
Terminated
2012-01-13
Reason
Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.
Root cause
Employee error
Action
Baxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.
Quantity
2 units
Distribution
Nationwide distribution: Puerto Rico.
Lot, serial or model codes
Product Code: 2M8161, Serial Numbers 15051625CC, 15020263CC
510(k) numbers
["K063696"]
PMA numbers
not on file
Product code
FRN
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register