Skip to the content

FDA Medical Device Recalls (openFDA)

Cytocell Ltd.: Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Recall number
Z-0741-2022
Recalling firm
Cytocell Ltd.
Product
Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024
Status
Terminated
Initiated
2022-01-18
Posted
not on file
Terminated
2024-11-08
Reason
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result
Root cause
Software Manufacturing/Software Deployment
Action
Cytocell issued Urgent Medical Recall Reference: VC/2022/001 letter on 1/18/22 to Distributors and End Users via email. Letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that you destroy the remaining inventory. We also suggest that laboratories undertake a review of any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of the probe being manufactured with the fluorophores the incorrect way around. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Quantity
19 kits
Distribution
US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.
Lot, serial or model codes
Lot Number: 075711 Exp. Date: 2023-06-22 UDI: (01)05055844901551(17)230622(10)075 711
510(k) numbers
not on file
PMA numbers
not on file
Product code
QDI
Firm FEI
3002646096
Firm city
Cambridge
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register