Skip to the content

FDA Medical Device Recalls (openFDA)

Horiba Instruments Inc: ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).

Recall number
Z-0741-2016
Recalling firm
Horiba Instruments Inc
Product
ABX PENTRA Lactic Acid CP ref. A11A01721 reagent is intended for the quantitative in vitro diagnostic determination of lactic acid in plasma by colorimetry. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).
Status
Terminated
Initiated
2016-01-11
Posted
2016-02-02
Terminated
2016-06-07
Reason
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Root cause
Under Investigation by firm
Action
Customer notification letters were sent on 01/11/16. Customers were provided the following instructions/actions: When performing tests that use the Tinder reaction, the blood draw should be done before treatment is given to the patient. Reagent inserts for the concerned HORIBA Medical Products have been updated and now include the following additional warning: Patients treated with N-Acetylcysteine (NAC) for Paracetamol overdose may generate a false low result. The letter request customers share the information with laboratory staff and retain the notification as part of their laboratory documentation. The letter states it is mandatory for customers to complete and return the enclosed response form within 10 days in order for the firm to verify the understanding of the notice. If any questions contact the local HORIBA Medical representative.
Quantity
216
Distribution
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
Lot, serial or model codes
All Lots
510(k) numbers
not on file
PMA numbers
not on file
Product code
KHP
Firm FEI
3002698356
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register