Skip to the content

FDA Medical Device Recalls (openFDA)

Hamilton Medical Inc: HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Recall number
Z-0741-2013
Recalling firm
Hamilton Medical Inc
Product
HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics
Status
Terminated
Initiated
2012-12-19
Posted
2013-01-31
Terminated
2014-04-22
Reason
Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.
Root cause
Software design
Action
Hamilton sent a Field Safety alert letter dated January 11, 2013 to all customers. The letter identified the affected product, explained the potential problem and a included Medical Device Safety Alert and Corrective Action document. Hamilton is working on a new software version for this issue. Customers are asked to update their operator manual with the update oxygen consumption formula page which provided revised information for calculating oxygen capacity. When the new software is available, Hamilton will provide loaner devices to each location while their unit is being upgraded. Questions should be addressed to Hamilton Medical at 800-426-6331 ext 215
Quantity
97 total ventilators; 69 of part number 161005 and 28 of part number 161006
Distribution
US Nationwide Distribution including the country of Switzerland.
Lot, serial or model codes
Software Versions 1.1.2 and below; Part numbers 161005, 161006.
510(k) numbers
["K120670"]
PMA numbers
not on file
Product code
CBK
Firm FEI
2937708
Firm city
Reno
Firm state
NV
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register