FDA Medical Device Recalls (openFDA)

Advanced Sterilization Products: STERRAD 200 Sterilizer, Hydrogen Peroxide Gas Plasma Sterilizer, Chemical Sterilizer. STERRAD 200; Product Usage: The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Recall number
Z-0741-2012
Recalling firm
Advanced Sterilization Products
Product
STERRAD 200 Sterilizer, Hydrogen Peroxide Gas Plasma Sterilizer, Chemical Sterilizer. STERRAD 200; Product Usage: The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Status
Terminated
Initiated
2012-01-23
Posted
2012-01-13
Terminated
2013-09-23
Reason
ASP determined that an internal filter in the vacuum pump on the STERRAD 200 System is subject to wear over time.
Root cause
Nonconforming Material/Component
Action
ASP will send an Urgent Medical Device Correction letter to all affected customers on January 23rd, 2012. The letter describes the product, problem and the actions to be taken. The letter informs customers that ASP has determined that the vacuum pump in the STERRAD 200 System may cause the emission of a slightly increased level of hydrogen peroxide (H2O2) vapor into the surrounding environment if not regularly serviced or replaced. The letter instructed customers to read the "Issue Description" and "Recommendations" sections in the letter, pass on the notice with any staff that work with or around the STERRAD 200, and maintain awareness of this communication until ASP has serviced the system. Also, a list of FAQ was provided. The letter states that ASP will replace, free of charge, the vacuum pumps on all STERRAD 200 Systems that have exceeded the pump manufacturer's recommended service interval. ASP has contracted with Stericycle to manage this field correction. Telephone numbers are provided in the letter: for questions related to this field correction, please contact Stericycle at 1-877-225-9750 and to report any complaints or suspected problems with the System, please contact ASP Professional Services directly at 1-888-783-7723, option # 2.
Quantity
395 units
Distribution
Worldwide Distribution - USA (nationwide)
Lot, serial or model codes
Product Code 10201
510(k) numbers
["K030429"]
PMA numbers
not on file
Product code
MLR
Firm FEI
3003662624
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register