Skip to the content

FDA Medical Device Recalls (openFDA)

Ortho Development Corporation: U-Joint Hex Head Driver 3.5mm, Part Number 228-0016, for the Escalade Acetabular Cup System. Driver is included in Part Number 270-9004-Escalade Acetabular Prep Kit and Part Number 228-9001-Triplus Acetabular Instrument Kit. The Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Hex Head Drive is used to place the screw that affixes the acetabular cup to the hip acetabulum.

Recall number
Z-0740-2012
Recalling firm
Ortho Development Corporation
Product
U-Joint Hex Head Driver 3.5mm, Part Number 228-0016, for the Escalade Acetabular Cup System. Driver is included in Part Number 270-9004-Escalade Acetabular Prep Kit and Part Number 228-9001-Triplus Acetabular Instrument Kit. The Acetabular Cup System is indicated for use in total hip arthroplasty procedures where the means of cup fixation is uncemented, biological fixation. The Hex Head Drive is used to place the screw that affixes the acetabular cup to the hip acetabulum.
Status
Terminated
Initiated
2011-12-19
Posted
2012-01-13
Terminated
2012-08-27
Reason
Tip of hex-head screwdriver may break off during use in hip replacement procedure.
Root cause
Device Design
Action
Ortho Development Corporation sent an "URGENT!! DEVICE RECALL" letter dated December 19, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to check to see if they have the affected device. Customers who have the device are to return the device to the firm, immediately. The firm will provide suitable replacements. Contact the Regulatory Affairs Manager for questions regarding this notice.
Quantity
37 units
Distribution
Nationwide Distribution-including the states of AZ, CA, FL, MI, OR, TX, and UT.
Lot, serial or model codes
Lot numbers 1365304, 1365305.
510(k) numbers
["K103384"]
PMA numbers
not on file
Product code
LZO
Firm FEI
3017298928
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register