Skip to the content

FDA Medical Device Recalls (openFDA)

NeuMoDx Molecular Inc: NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.

Recall number
Z-0738-2022
Recalling firm
NeuMoDx Molecular Inc
Product
NeuMoDx Flu A-B/RSV/SARS-CoV-2 Vantage Test Strip, Ref # 300900 (marketed under EUA202947). Distributed in boxes of 6 assay strips in individual pouches. Box, pouch and strips all labeled with product name, ref #, lot #, and expiration date.
Status
Open, Classified
Initiated
2022-01-21
Posted
not on file
Terminated
not on file
Reason
A risk for product contamination was identified for the affected product lot. If used, there is an increased potential for SARS-CoV-2 false positive results when using the product.
Root cause
Under Investigation by firm
Action
Customers were contacted by phone 21 Jan 2022 and sent a follow up email with instructions to stop using Lot 116305 S/L02, discard remaining inventory, and review all SARS-CoV-2 positive results previously obtained using the affected product lot to exclude erroneous diagnosis and treatment. Customers are to forward this information to all individuals and departments within their organization using the affected product. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com.
Quantity
102 boxes
Distribution
US Nationwide distribution in the states of FL, MD, MI, NY.
Lot, serial or model codes
Lot: 116305 S/L-02 (Sublot 02),UDI: 10814278020830
510(k) numbers
not on file
PMA numbers
not on file
Product code
QJR
Firm FEI
3010729479
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register