Skip to the content

FDA Medical Device Recalls (openFDA)

Diagnostica Stago, Inc.: STA SYSTEM CONROL N + P (ref. 00678)

Recall number
Z-0738-2020
Recalling firm
Diagnostica Stago, Inc.
Product
STA SYSTEM CONROL N + P (ref. 00678)
Status
Terminated
Initiated
2019-11-18
Posted
2019-12-23
Terminated
2023-01-12
Reason
A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.
Root cause
Process control
Action
On November 18, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: FIELD SAFETY NOTICE Combined use of the following reagents STA UNICALIBRATOR (ref. 00675) and STA DEFICIENT VIII (ref. 00725) STA DEFICIENT IX (ref. 00724) Dear Customer, You are using the STA Unicalibrator (00675) and STA Deficient VIII (00725) and/or STA Deficient IX (00724) and according to our records, you have ordered and received in your laboratory one or several kits of the STA - Unicalibrator listed in Appendix 1. Following the identification of a defect, please see the information below about the reagents and lots listed in Appendix 1. Description: Following a customer complaint, Stago has investigated and confirmed a positive bias on some STA - Unicalibrator levels. These high levels affect tests using STA - Deficient IX and STA - Deficient VIII reagents. Stago has accordingly reduced the concerned Unicalibrator plasma levels (see new levels on the attached lot-specific barcode sheets). According to our internal testing results, the reduced Unicalibrator plasma levels result in a decrease of Factor VIII and IX results on normal and below-normal patient plasmas. This change has no impact on the classification of hemophiliac patients (type A or B) as severe, moderate or mild. The reassignment of the Unicalibrator plasma levels necessitated a change in the acceptance ranges on the corresponding Quality Control plasmas for Factor VIII and IX parameters with the STA - Deficient VIII and IX reagents. You will find new barcode sheets for each lot of the affected reagents in the attachments. Actions: Upon receipt of this letter, we kindly ask you to: - No longer use the values indicated for STA - Deficient VIII and IX reagents on the barcode sheets in your possession (for the reagents & lots listed in Appendix 1). - Use the new barcode sheets (attached): - For STA - Unicalibrator: Use the analyzer-spec
Quantity
5,437
Distribution
US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK
Lot, serial or model codes
REF: 00678 Lot Numbers: 253145, 253416, 253655, 254040, 254206, 254401, 254573, 254760  UDIs: (01)03607450006780(11)171130(17)191130(10)253145(241)00678 (01)03607450006780(11)180131(17)200131(10)253416(241)00678 (01)03607450006780(11)180331(17)200331(10)253655(241)00678 (01)03607450006780(11)180630(17)200630(10)254040(241)00678 (01)03607450006780(11)180731(17)200731(10)254206(241)00678 (01)03607450006780(11)180930(17)200930(10)254401(241)00678 (01)03607450006780(11)181031(17)201031(10)254573(241)00678 (01)03607450006780(11)181130(17)201130(10)254760(241)00678
510(k) numbers
["K943518"]
PMA numbers
not on file
Product code
GGN
Firm FEI
2245451
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.