Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation: V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001

Recall number
Z-0737-2024
Recalling firm
Steris Corporation
Product
V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001
Status
Open, Classified
Initiated
2023-11-16
Posted
2024-01-17
Terminated
not on file
Reason
Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
Root cause
Under Investigation by firm
Action
Steris issued Urgent Medical Device Recall Letter to Distributors and Customers on November 16, 2023. Letter states reason for recall, health risk and action to take: Please ensure the following steps are completed: 1. Please immediately inspect your on-hand inventory for product affected by this recall. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. STERIS will coordinate credit issuance upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: Regulatory_Compliance@STERIS.com or via fax to (440) 392-8963. If you have any questions regarding this recall, please contact regulatory_Compliance@STERIS.com.
Quantity
4,214 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 10885403065866 Lot Numbers: L22XRD, C23XRD, E23XRD, and F23XRD.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRW
Firm FEI
1527821
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register