Skip to the content

FDA Medical Device Recalls (openFDA)

Bard Peripheral Vascular Inc: Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters.

Recall number
Z-0737-2016
Recalling firm
Bard Peripheral Vascular Inc
Product
Denali Filter-Jugular/Subclavian DL950J Denali Filter-Femoral Delivery DL950F. Intravascular filters.
Status
Terminated
Initiated
2015-03-13
Posted
2016-02-01
Terminated
2016-02-01
Reason
IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys.
Root cause
Labeling Change Control
Action
The firm sent out Medical Device Recall customer notification letters on 03/13/15. The notification letter stated that there is no required action on the customer's part. Customers were told that if the device did not meet their needs then the firm will provide a replacement unit. Effective immediately the firm is implementing a change in the ordering code for the device to accurately reflect the approved information with the IFU and corresponding retrieval kit. Any questions the firm states to call Bard Canada: 1-800-268-2862 Western Canada 1-800-387-9473 Ontario 1-800-387-7851 Eastern Canada
Quantity
1,183 units
Distribution
Distributed to CA only.
Lot, serial or model codes
Product Code: DL950J Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699  Product Code: DL950F Lot Number: GFYA3093, GFYC3745, GFYD2734, GFYF3878, GFYF3879, GFYG3877, GFYG3880, GFYI2699
510(k) numbers
["K130366"]
PMA numbers
not on file
Product code
DTK
Firm FEI
2020394
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register