Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 209 Inc.: NicVue Software version 2.9.2 and 3.0.1. Medical Device Data System (MDDS) Intended use: The NicVue software is intended to be the patient database for various Nicolet applications. The NicVue software is intended to aid the hospital in tracking patient data from collection to review to archive. The NicVue Software includes the following features and intended uses: Launch Applications, Enter patient information, Import/Export patient information from the Hospital Information System, Move data files, Archive data files on DVD or CDROM.

Recall number
Z-0737-2012
Recalling firm
CareFusion 209 Inc.
Product
NicVue Software version 2.9.2 and 3.0.1. Medical Device Data System (MDDS) Intended use: The NicVue software is intended to be the patient database for various Nicolet applications. The NicVue software is intended to aid the hospital in tracking patient data from collection to review to archive. The NicVue Software includes the following features and intended uses: Launch Applications, Enter patient information, Import/Export patient information from the Hospital Information System, Move data files, Archive data files on DVD or CDROM.
Status
Terminated
Initiated
2011-12-16
Posted
2012-01-13
Terminated
2012-10-11
Reason
Under certain work flow processes, when selecting a new patient name in NicVue, the associated exams list may not properly refresh, and may continue to display the previous patient's exams.
Root cause
Software design
Action
The firm, CareFusion, sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated December 16, 2011 to its customers. The letter included a "Description of the issue" and "Actions you must take to correct this issue". The customers were instructed to follow the instructions included with the CD to update all of your NicVue systems; complete and return the SOFTWARE UPDATE VERIFICATION FORM to CareFusion-NeuroCare via fax to: +1 608 829 8517 ; mail to ATTN: Regulatory Affairs, CareFusion-NeuroCare, 1850 Deming Way, Middleton, WI USA 53562 and/or email (PDF) to: Susan.Niesen@CareFusion.com, and if you have transferred any of these products to another location, send those users a copy of this Field Correction notification and notify CareFusion-NeuroCare at +1 800 356 0007 or +1 608 829 8500 ext. 5129, or via email at susan.niesen@carefusion.com of this transfer. If assistance is needed in completing these actions, contact CareFusion NeuroCare Call Center at +1 800 356 0007 or +1 608 829 8500 for further directions; press option 2 for Technical Support.
Quantity
1,676 ( 1062 OUS, 614 US)
Distribution
Worldwide distribution: USA (nationwide) including states of: AL, AK, AZ, AR, CA, CO, DE, DC, FL, GA, HI, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA,WV, and WI; and countries of: ARGENTINA, AUSTRALIA, BULGARIA, BANGLADESH, BELGIUM, CHILE COLOMBIA, CYPRUS, CZECH REPUBLIC, CANADA, CHINA, CROATIA, CURACO, DENMARK, DOMINICANA, EGYPT, FRANCE,GERMANY, HUNGARY, HONG KONG, IRELAND, INDIA, ITALY, JAPAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NORWAY, NETHERLANDS, OMAN, PARAGUAY, PERU, POLAND, PORTUGAL, PALESTINIAN, PAKISTAN, PHILIPPINES,QATAR, ROMANIA, SAUDI ARABIA, SINGAPORE, SERBIA, SLOVENIA, SPAIN, SLOVAKIA, SWITZERLAND, SRI LANKA, RUSSIA, TAIWAN, THAILAND, TURKEY, UNITED KINGDOM, URUGUAY, UNITED ARB EMIRATES, and VIET NAM.
Lot, serial or model codes
Part Numbers:   828-050000, 828-062400, 828-062500, 828-062600, 828-050100, 828-062900, 828-063100, 828-063200, 828-063300, 828-063500
510(k) numbers
not on file
PMA numbers
not on file
Product code
OUG
Firm FEI
3010611950
Firm city
Middleton
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register