Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Lotus Edge Valve System, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System, 27mm, sterile, REF H749LVS270, GTIN 08714729940838, Made in Ireland, Ballybrit Business Park, Galway, Ireland.

Recall number
Z-0736-2021
Recalling firm
Boston Scientific Corporation
Product
Lotus Edge Valve System, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System, 27mm, sterile, REF H749LVS270, GTIN 08714729940838, Made in Ireland, Ballybrit Business Park, Galway, Ireland.
Status
Terminated
Initiated
2020-11-17
Posted
2021-01-09
Terminated
2022-01-24
Reason
Failure to execute the visual inspection correctly prior to locking the valve during the deployment process can lead to the inability to release the delivery system.
Root cause
Device Design
Action
The recalling firm issued a press release on 11/17/2020, as well as letters dated 11/17/2020 issued via overnight mail delivery to the U.S. direct account medical facilities and clinical investigators. OUS customers were also issued the letter on 11/17/2020 through various methods, including email, courier, hand-delivery, and phone. The letter was flagged "Medical Device Recall-Immediate Action Required" and said BSC was conducting a removal of the unused inventory and there was no safety issue for patients who previously received the valve. The letter explained the details surrounding the reason for removal from the field. Regarding actions to be taken by the customer, BSC recommended that patients who have the device continue with their routine follow-up care and no additional action is needed for them. Further distribution or use of the remaining product was to be ceased immediately. The product was to be immediately segregated and returned to Boston Scientific per their enclosed instructions. If the customer was a distributor, they were instructed to forward the notification to their customers as the recall was to the hospital level. A Reply Verification Tracking Form was enclosed to be completed and returned via email or FAX even if the customer no longer has any of the devices.
Quantity
386 devices
Distribution
Distribution was made to AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. Government distribution was also made. Foreign distribution was made to Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, South Korea, Spain, Sweden, and Switzerland.
Lot, serial or model codes
GTIN 08714729940838, All unexpired lot numbers, expiration dates 2-Nov-2020 through 29-Oct-2021
510(k) numbers
not on file
PMA numbers
["P180029"]
Product code
NPT
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register