Skip to the content

FDA Medical Device Recalls (openFDA)

Biomerieux, Inc.: NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).

Recall number
Z-0735-04
Recalling firm
Biomerieux, Inc.
Product
NucliSensEasy Q Incubator, Product Number 285204 (this number can be found at the back of the instrument).
Status
Terminated
Initiated
2004-02-26
Posted
2004-07-20
Terminated
2004-03-24
Reason
Instrument malfunction. Electrical short circuit may occur resulting in damage or a fire hazard.
Root cause
Other
Action
This correction is being handled as an instrument repair (rework) by bioMerieux Field Service; therefore no communication has been sent to the customer sites. The repair involves heat sink replacement or use of insulating tape on the rear side of the keypad.
Quantity
3 units
Distribution
CT, MD, MT
Lot, serial or model codes
Catalog (or Product) number 285204, Serial Number 36061, 36065 and 36068
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJQ
Firm FEI
1039284
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register