FDA Medical Device Recalls (openFDA)
Accuray Incorporated: CyberKnife Robotic Radiosurgery System using software version 10.6; Catalog/Part Number: 54000 UDI: M658053301 0 Radiology: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Recall number
- Z-0734-2017
- Recalling firm
- Accuray Incorporated
- Product
- CyberKnife Robotic Radiosurgery System using software version 10.6; Catalog/Part Number: 54000 UDI: M658053301 0 Radiology: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Status
- Terminated
- Initiated
- 2016-11-22
- Posted
- 2016-12-12
- Terminated
- 2017-05-22
- Reason
- Accuray has become aware of a potential safety issue involving unexpected treatment robot motion when removing an accessory. This issue is limited to certain CyberKnife Systems with software version 1 0.6.
- Root cause
- Software design
- Action
- Accuray sent an Urgent Safety Advisory Notification letter dated November 22, 2016, to all affected customers. Letters alerted users to the issue and the sequence of events necessary for it to occur. Accuray reps will visit all sites to perform a software upgrade to correct the issue. For more information or any questions about this notification, please contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solution Center at http://www.accuray.com/services-support. For further questions, please call (408) 716-4600.
- Quantity
- 40 affected devices
- Distribution
- Worldwide Distribution - US Distribution to the states of : PA, WA, MT, NY, SC, CA, MN, AK, NJ, FL, KY, WI., and to the countries of : Bulgaria, Canada, France, Germany, India, Japan, Luxembourg, Netherlands, Portugal, Switzerland and Turkey
- Lot, serial or model codes
- Catalog/Part Number: 54000 manufacturing or serial number of UDI: M658053301 0; Serial numbers: C0330 C0334 C0336 C0338 C0342 C0344 C0348 C0349 C0351 C0358 C0359 C0360 C0361 C0363 C0364 C0366 C0367 C0370 C0372 C0376 C0387 C0389 C0390 C0391 C0393 C0394 C0395 C0396 C0397 C0398 C0400 C0402 C0404 C0406 C0407 C0408 C0409 C0410 C0413 C0414
- 510(k) numbers
- ["K122137"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1000123585
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28