Skip to the content

FDA Medical Device Recalls (openFDA)

Johnson & Johnson Vision Care, Inc.: Outer Carton labeled in part:" 1-DAY ACUVUE MOIST BRAND CONTACT LENSES WITH LACREON FOR ASTIGMATISM***Contact Lenses for Daily Wear***D -1.50***-1.25 100***2016/07***BC 8.5***DIA 14.5***Made in USA***LOT B00BVTN0VS***" Individual Lenses labeled in part: "***1-DAY ACUVUE MOIST FOR ASTIGMATISM WITH LACREON***BC 8.5***DIA 14.5***AXIS -1.75 -0.75***080***2016/06***LOT B00BSBH***" It is intended for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with not diseased eyes who may have 1.00D or less of astigmatism.

Recall number
Z-0734-2012
Recalling firm
Johnson & Johnson Vision Care, Inc.
Product
Outer Carton labeled in part:" 1-DAY ACUVUE MOIST BRAND CONTACT LENSES WITH LACREON FOR ASTIGMATISM***Contact Lenses for Daily Wear***D -1.50***-1.25 100***2016/07***BC 8.5***DIA 14.5***Made in USA***LOT B00BVTN0VS***" Individual Lenses labeled in part: "***1-DAY ACUVUE MOIST FOR ASTIGMATISM WITH LACREON***BC 8.5***DIA 14.5***AXIS -1.75 -0.75***080***2016/06***LOT B00BSBH***" It is intended for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with not diseased eyes who may have 1.00D or less of astigmatism.
Status
Terminated
Initiated
2011-12-01
Posted
2012-01-12
Terminated
2012-06-18
Reason
Johnson & Johnson Vision Care, Inc. recalled their 1-DAY ACUVUE MOIST Brand Contact Lenses for ASTIGMATISM due to a lot being mixed. The secondary carton lot number and power are not the same as on the primary package inside the carton.
Root cause
Employee error
Action
The firm, Johnson & Johnson Vision Care, sent an "IMPORTANT PRODUCT NOTIFICATION" letter dated 1 December 2011 to all consignees/customers. The letter describes the product, problem and actions to be taken. The customers were instructed to check their practice/store inventory for product affected by this discrepancy in labeling; if you dispensed this product to a patient; contact the patient and replace lenses to all intended visual correction and performance with the correct lens power; complete and return ACKNOWLEDGEMENT OF RECEIPT Form via fax to: 904-443-3442 or email: VPIWEB@visus.jn.com and return any cartons labeled with the noted lot number to Customer Relations. Johnson & Johnson will replace them with the correct product. For any questions regarding this labeling situation, contact Customer Relations at 1-800-843-2020.
Quantity
152 cartons
Distribution
Worldwide distribution: USA (nationwide) including states of: NY and FL and countries of: Finland, France, Germany, Italy, Netherlands, Norway, Denmark, Switzerland, Sweden, and the United Kingdom.
Lot, serial or model codes
External carton Lot Number B00BVTN -1.50 -1.25cyl. 100¿ (30 primary packages per secondary carton Internal package Lot Number B00BSBH -1.75 -0.75cyl. 80¿ (30 primary packages per secondary carton)
510(k) numbers
["K051900"]
PMA numbers
not on file
Product code
MVN
Firm FEI
1000222023
Firm city
Jacksonville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register