Waldemar Link GmbH & Co. KG (Mfg Site): LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.

Recall number
Z-0733-2019
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Product
LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of loosened femoral prosthesis components involving extensive bone resorption of the proximal femur and widening of the medullary cavity or marked thinning of proximal femoral cortical bone Revision of loosened femoral prosthesis components with periprosthetic/subprosthetic fracture Deformed proximal femur due to fractures or osteotomies Correction of bone deficiencies, e.g. due to tumors Large post-revision and post-trauma segmental bone defects Oncological and revision surgery from tibial to hip area (in conjunction with Endo-Model SL Rotational and Hinge Knee Prostheses) The device is intended for cemented and cementless use.
Status
Terminated
Initiated
2018-12-07
Posted
not on file
Terminated
2019-12-17
Reason
The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
Root cause
Under Investigation by firm
Action
Urgent Medical Device Recall notification letters dated 12/7/18 were distributed to customers. Please identify your total inventory for this product. Please take the following actions: 1) Immediately cease use and segregate the above described product. 2) Complete the enclosed Consignee Reply Form, even if you have no inventory. 3) Return the above product and a copy of the Consignee Reply Form using Federal Express Account #: 007183879 to: LinkBio Corporation 101 Roundhill Drive Rockaway, NJ 07446 ATT: Nikhil Mangale Recall Return Authorization #: R-2018-03
Quantity
11
Distribution
US Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
Lot, serial or model codes
Lot numbers: 1435077 1515029 1718103 1435077 150401/0217 1734168 1726056 1718103 1734168 1515029
510(k) numbers
["K142187"]
PMA numbers
not on file
Product code
LZO
Firm FEI
3003386935
Firm city
Norderstedt
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register