Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: Cook Spectrum Minocycline/Rifampin Impregnated Double Lumen Central Venous Catheter Tray Cook Spectrum Minocycline/Rifampin Impregnated Triple Lumen Central Venous Catheter Tray

Recall number
Z-0733-2017
Recalling firm
Cook Inc.
Product
Cook Spectrum Minocycline/Rifampin Impregnated Double Lumen Central Venous Catheter Tray Cook Spectrum Minocycline/Rifampin Impregnated Triple Lumen Central Venous Catheter Tray
Status
Terminated
Initiated
2016-11-28
Posted
2016-12-12
Terminated
2017-09-06
Reason
Cook Medical has received reports of the trays containing the incorrect needle, which in some cases could result in the inability to pass a wire guide through the needle. Potential adverse events that may occur as a result of utilizing needles of incorrect size include delay in treatment, which could be critical in an emergent situation. There have been no reports of illness or injury associated with this issue.
Root cause
Mixed-up of materials/components
Action
Cook Medical sent an Urgent - Medical Device Recall letter dated November 23, 2016. to all affected customers. The letter identified the product the problem and the action needed to be taken. Action to be taken: 1. Please review the attached list of affected products and lot numbers that were shipped to your account and quarantine any affected products that remain unused. 2. Immediately collect and return all unused affected products to Cook Medical as soon as possible for credit. 3. Please complete the attached Acknowledgement and Receipt Form and return it to Cook Medical. 4. Please report any adverse event to Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235, Monday through Friday between 7:30 am and 5:00 pm Eastern Standard Time or e-mail at fieldactionsna@cookmedical.com. Transmission of this notice: This notice must be passed on to appropriate personnel, including down to the user level, within your organization or to any organization where the potentially affected devices have been transferred. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at (800) 457-4500 or 1(812) 339-2235.
Quantity
99 units
Distribution
AZ CA DC GA IL LA MN MS NC NV NY OK PA TN TX WI
Lot, serial or model codes
catalog number: C-UDLMY-401J-ABRM-HC-FST Lot 6498570 catalog number: C-UTLMY-501J-ABRM-HC-FST Lot 6501835  Exp date 9/30/2018
510(k) numbers
["K033843"]
PMA numbers
not on file
Product code
FOZ
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register