Philips North America: (1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436

Recall number
Z-0732-2024
Recalling firm
Philips North America
Product
(1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436
Status
Open, Classified
Initiated
2023-11-28
Posted
2024-01-13
Terminated
not on file
Reason
Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
Root cause
Software design
Action
On November 28, 2023, the firm notified affected consignees through URGENT Medical Device Correction letters. This failure is applicable to Apple iOS devices only. If customers use iOS devices (e.g. iPhone, iPad, etc.) to receive event notifications: 1) Please conduct patient monitoring at the central station or patient bedside. 2) Please do not use iOS event notifications for patient monitoring. A Philips representative will contact you to schedule a visit to install a software patch to your device which will correct the current issue. Patch PIC iX 4.2.2 will be provided for PIC iX 4.X and patch CareEvent C.03.07 will be provided for CareEvent C.03.X. If you need any further information, please contact your local Philips representative, or call 1-800-722- 9377.
Quantity
254 US; 37 OUS
Distribution
Worldwide distribution: US (Nationwide) to states of: CA, GA, IA, MA, MI, MN, NJ, NM, NY, OH, PA, TN, TX, WA, WV; and OUS (Foreign) distribution to countries of: Australia, Austria, Belarus, Belgium, Brazil, Canada, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, India, Indonesia, Ireland, Italy, Japan, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd.Arab.Emir., and Uzbekistan
Lot, serial or model codes
UDI-DI: 00884838099128 Software: CareEvent Version C.03.X
510(k) numbers
["K153702","K161164"]
PMA numbers
not on file
Product code
MHX
Firm FEI
3006648320
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register