Skip to the content

FDA Medical Device Recalls (openFDA)

Vital Diagnostics, Inc.: Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum

Recall number
Z-0732-2014
Recalling firm
Vital Diagnostics, Inc.
Product
Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum
Status
Terminated
Initiated
2013-08-29
Posted
2014-01-14
Terminated
2015-08-26
Reason
High positive calcium bias on plasma sample versus results on serum
Root cause
Device Design
Action
Vital Diagnostics issued a Recall Notice dated August 29, 2013, to all affected customers via certified mail. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use plasma for the determination of total calcium when using the affected product. The use of serum is acceptable, thus the product can still be used to determine total calcium with this sample type. Vital Diagnostics released a corection to the package insert for the Envoy Calcium Reagent. Customers with the affected product in their laboratory, were instructed to discard the insert packed with the product and replace it with the insert enclosed with this bulletin. If product was transferred to another location, the bulletin should be forwarded to that location. The package insert was revised indicating to only use Serum samples when running Calcium. Vital Diagnostics issued a revised (expanded) notice dated September 19, 2013, to include the Consult Diagnostics Eon Calcium Reagent (PN: 1429) requesting users to discontinue using the single vial Eon Calcium kit (PN: 1429). Customers with questions were instructed to call Vital Diagnostics Technical Support at 1-855-354-8324.. For questions regarding this recall call 401-642-8400.
Quantity
395 units
Distribution
Nationwide Distribution
Lot, serial or model codes
All lots
510(k) numbers
["K120626"]
PMA numbers
not on file
Product code
CJY
Firm FEI
1717966
Firm city
Lincoln
Firm state
RI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register