Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical Solutions, Malvern, PA 19355

Recall number
Z-0732-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical Solutions, Malvern, PA 19355
Status
Terminated
Initiated
2007-11-27
Posted
2008-02-23
Terminated
2008-09-17
Reason
Inconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)
Root cause
Software design
Action
The recalling firm issued a Customer Safety Advisory, via certified mail. to the affected customers on 11/27/07. The letter informs the customer of the issue and also provides instruction to detect and avoid its occurrence. In addition, the firm is developing a software update to correct the issue, which will be installed on the affected systems when it is released.
Quantity
10 units
Distribution
Nationwide Distribution, including states of AL, MO, NC, NY, OH, OR, TN, and TX.
Lot, serial or model codes
Serial Numbers: 1005, 1065, 1076, 1112, 1202, 1242, 1269, 1278, 1279, and 1282.
510(k) numbers
["K071114"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register