Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: MOSAIQ Oncology Information System The product is used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be use wherever radiotherapy and chemotherapy are prescribed.

Recall number
Z-0731-2017
Recalling firm
Elekta, Inc.
Product
MOSAIQ Oncology Information System The product is used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be use wherever radiotherapy and chemotherapy are prescribed.
Status
Terminated
Initiated
2016-12-02
Posted
not on file
Terminated
2021-11-05
Reason
Edits to Particle field definition parameters may not be saved when the field definition window is saved.
Root cause
Software design
Action
The firm, Elekta, sent an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter dated 11/2016 was sent to affected customers on 12/2/2016. The letter describes the product, problem and actions to be taken. The customers were instructed to save and re-open the field definition form when manual edits have been performed to confirm that the edits were save; post this notice in a place accessible to all users; advise the appropriate personnel working with the product the content of the letter; and complete and return the Acknowledgement Form via fax to: 408 830 8003 immediately upon receipt but no later than 30 days. Elekta will resolve the problem with the MOSAIQ 2.60 SP7F, MOSAIQ 2.60 SP14 and 2.64 SP3. Customers are encouraged to contact Elekta support to get the corrected version. If you have any questions, contact RA Manager Americas at 1-770-300-9725 or email to: Elekta_Global_PMS@elekta.com.
Quantity
47
Distribution
Worldwide Distribution-US (nationwide) to states of: FL, IL, MO, NJ, OH, OK, TN, TX, VA, and countries of: Italy, Japan, South Korea, Sweden and United Kingdom.
Lot, serial or model codes
Versions 2.60 an higher
510(k) numbers
["K141572"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Atlanta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register