FDA Medical Device Recalls (openFDA)
Merge Healthcare, Inc.: Merge Cardio software. Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, hemodynamic studies and reports, measurements (via import from DICOM Structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data
- Recall number
- Z-0730-2017
- Recalling firm
- Merge Healthcare, Inc.
- Product
- Merge Cardio software. Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including images, hemodynamic studies and reports, measurements (via import from DICOM Structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data
- Status
- Terminated
- Initiated
- 2015-10-01
- Posted
- not on file
- Terminated
- 2018-03-23
- Reason
- Reporting feature times out after inactivity for more than an hour sending the user back to the study list, which causes all reporting data being entered to be lost.
- Root cause
- Software design
- Action
- MERGE sent an Urgent Medical Device Recall letter dated March 28, 2016, to all affected customers. Non-responders were issued a second letter dated July 11, 2016, The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to reply using the enclosed form and the return addressed envelope and also to ensure that all users of the product were provided with the notification. If product was further distributed, customer were asked to identify their customers and notify them at once of the product recall. Customers with questions were advised to email recall@merge.com. For questions regarding this recall call 877-741-5369.
- Quantity
- 116 sites potentially have the effected versions
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Versions V10.1 and 10.1.1
- 510(k) numbers
- ["K051649"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 1000116130
- Firm city
- Hartland
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28