Skip to the content

FDA Medical Device Recalls (openFDA)

O&M HALYARD, INC.: O&M Halyard SOFT TOUCH II Surgical Mask, Blue-Disposable surgical mask Product Code: 47500

Recall number
Z-0729-2025
Recalling firm
O&M HALYARD, INC.
Product
O&M Halyard SOFT TOUCH II Surgical Mask, Blue-Disposable surgical mask Product Code: 47500
Status
Open, Classified
Initiated
2024-11-08
Posted
2024-12-20
Terminated
not on file
Reason
Not meeting the bioburden testing requirements of EN 14683 required for masks sold in Europe. Face masks sold in the United States are not required to meet this standard. However, if the standard is printed on the packaging, compliance is required.
Root cause
Nonconforming Material/Component
Action
O&M Halyard issued Urgent: Product Field Action via email on 11/8/24 to Distributors and End-Users. Letter states reason for recall, health risk and action to take: Customers (end-users): 1. REVIEW your inventory for the affected product code and specific lot number. 2. DESTROY or DISCARD the product per the facility s standard disposal procedure for face masks. 3. COMMUNICATE this Field Action to staff to ensure awareness of it. 4. RETURN the completed Field Action Response Form to OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. 5. Any adverse events associated with the use of the affected product should be reported to complaints@owens-minor.com. 6. If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: 1. REVIEW your inventory for the affected product code and specific lot number. 2. DESTROY or DISCARD the product per the facility s standard disposal procedure for face masks. 3. COMMUNICATE this voluntary Field Action to staff to ensure awareness of it. 4. If any of the affected lot identified above was further distributed, please ensure all end-users are appropriately notified of this Field Action and maintain records of effectiveness. 5. Any adverse events associated with the use of the affected product should be reported to complaints@owens-minor.com. 6. RETURN the completed Field Action Response Form to OMRA_RECALLS@owens-minor.com as soon as possible even if you have no current product inventory. If you have any questions, please email OMRA_RECALLS@owens-minor.com using the Event # FA-2024-034 in the header of the response.
Quantity
637 cases
Distribution
Nationwide
Lot, serial or model codes
UDI-DI: 20680651475002 Lot Number: AM4119961 Exp. Date: 2029-04-28
510(k) numbers
["K823078","K831148"]
PMA numbers
not on file
Product code
FXX
Firm FEI
3014527898
Firm city
Mechanicsville
Firm state
VA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register