Cardinal Health Inc.: Cordis Precise Nitinol Stent System. The Cordis PRECISE Nitinol Stent System devices are designed to deliver a flexible, self-expanding endoluminal stent to the carotid vasculature via over-the-wire (OTW) 5.5F or 6F sheathed delivery systems. The stent is cut from a solid nitinol tube into a fine mesh ( Z configuration) design. The delivery systems consist mainly of an inner shaft and an outer sheath with radiopaque markers, and a Tuohy Borst valve. The inner shaft terminates distally in a catheter tip and originates proximally in a luer hub designed to accept a 0.018 guidewire. The delivery systems have a nominal working length of 135 cm. The self-expanding PRECISE stent is constrained within the space between the inner shaft and the outer sheath, located between distal and proximal stent makers on the inner shaft. The stent expands to its unconstrained diameter when released from the deployment catheter into the carotid artery. Upon deployment the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2mm smaller in diameter than the unconstrained diameter of the stent.

Recall number
Z-0729-2020
Recalling firm
Cardinal Health Inc.
Product
Cordis Precise Nitinol Stent System. The Cordis PRECISE Nitinol Stent System devices are designed to deliver a flexible, self-expanding endoluminal stent to the carotid vasculature via over-the-wire (OTW) 5.5F or 6F sheathed delivery systems. The stent is cut from a solid nitinol tube into a fine mesh ( Z configuration) design. The delivery systems consist mainly of an inner shaft and an outer sheath with radiopaque markers, and a Tuohy Borst valve. The inner shaft terminates distally in a catheter tip and originates proximally in a luer hub designed to accept a 0.018 guidewire. The delivery systems have a nominal working length of 135 cm. The self-expanding PRECISE stent is constrained within the space between the inner shaft and the outer sheath, located between distal and proximal stent makers on the inner shaft. The stent expands to its unconstrained diameter when released from the deployment catheter into the carotid artery. Upon deployment the stent forms an open lattice and pushes outward on the luminal surface, helping to maintain the patency of the artery. Due to the self-expanding behavior of nitinol, the stents are indicated for placement into vessels that are 1-2mm smaller in diameter than the unconstrained diameter of the stent.
Status
Open, Classified
Initiated
2019-11-14
Posted
not on file
Terminated
not on file
Reason
Cordis has initiated a labeling correction for S.M.A.R.T. and PRECISE stents due to incorrect MRI information on the IFU. The correct classification is "MRI Conditional" not "MRI Safe".
Root cause
Under Investigation by firm
Action
Customers were notified with a Medical Device Correction letter. The letter was sent via overnight delivery beginning 11/14/2019. The letter identified affected product, provided a description of the problem, and asked that a copy of the letter be retained where the product is stored. Customer were asked to completed and return an Acknowledgement Form. Questions can be directed to the Cordis Endovascular Medical Information Line 877-DEVICE5 (877-338-4235) Monday through Friday 9:00 am - 5:00 pm EST.
Quantity
267,247 units total (262,641 OUS, 4,606 US)
Distribution
Worldwide distribution - US Nationwide including Puerto Rico and countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaiian, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Repulic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Saint-Barthelemy, Saudia Arabia, Sinapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela.
Lot, serial or model codes
All Lot Codes. Affected SKUs: N1020SB, N1020SB, N1020SC, N1030SB, N1030SB, N1030SC, N1040SB, N1040SB, N1040SC, N520SB, N520SC,  N530SB, N530SC, N540SB, N540SC, N620SB, N620SC, N630SB, N630SC, N640SB, N640SC, N720SB, N720SC, N730SB, N730SC, N740SB, N740SC,  N820SB, N820SB, N820SC, N830SB, N830SB, N830SC, N840SB, N840SB, N840SC, N920SB, N920SB, N920SC, N930SB, N930SB, N930SC, N940SB, N940SB, N940SC, P05020RXB, P05030RXB, P05040RXB, P0520RXCE, P0530RXCE, P0540RXCE, P06020RXB, P06030RXB, P06040RXB, P0620RXCE, P0630RXCE, P0640RXCE, P07020RXB, P07030RXB, P07040RXB, P0720RXCE, P0730RXCE, P0740RXCE, P08020RXB, P08030RXB, P08040RXB, P0820RXCE, P0830RXCE, P0840RXCE, P09020RXB, P09030RXB, P0920RXCE, P0930RXCE, P0940RXCE, P10020RXB, P1020RXCE, P1030RXCE, P1040RXCE.
510(k) numbers
not on file
PMA numbers
["P030047"]
Product code
NIM
Firm FEI
3004008697
Firm city
Dublin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register