Skip to the content

FDA Medical Device Recalls (openFDA)

Perkinelmer Life Sciences, Inc.: JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.

Recall number
Z-0728-2018
Recalling firm
Perkinelmer Life Sciences, Inc.
Product
JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.
Status
Terminated
Initiated
2017-08-02
Posted
not on file
Terminated
2018-07-31
Reason
We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.
Root cause
Nonconforming Material/Component
Action
On August 4 2017 an URGENT Medical Device Correction letter titled "Horizontal Motion Error" was issued to customers informing them of the error notice and to follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. Question or concerns can be directed to Technical Support at 800-762-4000
Quantity
7
Distribution
Nationwide
Lot, serial or model codes
AJM4G01, AJL8001, AJI4G01, AJL8M01, CJM8M01, CJL8M01, CJL8001
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQW
Firm FEI
1000513616
Firm city
Boston
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register