Perkinelmer Life Sciences, Inc.: JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.
- Recall number
- Z-0728-2018
- Recalling firm
- Perkinelmer Life Sciences, Inc.
- Product
- JANUS & JANUS G3 Automated Workstation is an automated programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows.
- Status
- Terminated
- Initiated
- 2017-08-02
- Posted
- not on file
- Terminated
- 2018-07-31
- Reason
- We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.
- Root cause
- Nonconforming Material/Component
- Action
- On August 4 2017 an URGENT Medical Device Correction letter titled "Horizontal Motion Error" was issued to customers informing them of the error notice and to follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. Question or concerns can be directed to Technical Support at 800-762-4000
- Quantity
- 7
- Distribution
- Nationwide
- Lot, serial or model codes
- AJM4G01, AJL8001, AJI4G01, AJL8M01, CJM8M01, CJL8M01, CJL8001
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JQW
- Firm FEI
- 1000513616
- Firm city
- Boston
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28