Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203

Recall number
Z-0727-03
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ICON 9.5 Software used on the ICON workstations designed for use with the E.CAM Emission Imaging Computed Tomography Systems; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
Status
Terminated
Initiated
2003-02-10
Posted
2003-04-16
Terminated
2005-01-19
Reason
ICON 9.5 software does not apply flood corrections during whole body SPECT studies with multiple bed positions.
Root cause
Other
Action
Siemens sent Customer Advisory letters dated 2/10/03 to all ICON 9.5 customers with Whole Body SPECT functionality on integrated camera systems including Diacam, Multispect 2, Multispect 3 and e.cam camera systems aquired on an ICON workstation. The accounts were informed of the problem related to applying the flood correction matrix to Whole Body SPECT studies in ICON version 9.5 software, and were provided the follwoing workaround: To ensure that the flood correction is applied, the user should acquire multiple single bed positions with the General SPECT predefine, indexing the bed manually between each SPECT acquisition. The data can then be combined into a Whole Body SPECT in the Whole Body SPECT Display program for review. In addition, users were requested to routinely perform standard quality control procedures at the facility to ensure proper image quality. No software upgrade is planned.
Quantity
600 units
Distribution
Nationwide, United Arab Emirates, Argentina, Austria, Australia, Barbados, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Equador, Egypt, France, Germany, Great Britain, Georgia, Greece, Hong Kong, Israel, India, Italy, Jordan, Japan, Korea, Latvia, Mexico, Niger, Norway, New Zealand, Philippines, Pakistan, Poland, Russia, Saudi Arabia, SIngapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine and the Netherlands.
Lot, serial or model codes
All Diacam, Multispect 2, Multispect 3 and e.cam camera systems with ICON workstations with ICON software version 9.5 and the optional Whole Body SPECT software.
510(k) numbers
["K903315","K914350"]
PMA numbers
not on file
Product code
JWM
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register